Device Settings; Factory Settings - Dräger HFG 000 Series Technical Manual

Table of Contents

Advertisement

8

Device settings

Only trained and qualified personnel may change the device settings.
8.1

Factory settings

Different settings can be selected to meet customer requirements on order. The
setting can be checked and changed with the X-dock maintenance station or the
Dräger CC-Vision PC software.
Function
Fresh air calibration without password
Bump test without password
Operation signal
Switch off allowed
Bluetooth
(X-am 8000 only, if installed)
Capture range
LEL factor
ch4 (methane)
H
2
STEL
TWA
Alarm A1
Alarm A2
Symbol for type of danger
Expired bump test interval
Expired calibration interval
Power-save mode display
1) The set capture range can be read out on the gas detector and activated or deactivated. The
capture range is activated at the factory in measuring mode. The capture range is always
deactivated in calibration mode.
2) An LEL factor can be adapted to national regulations using the Dräger CC-Vision PC software.
|
Technical manual
X-am® 3500/8000
®
1)
2)
(hydrogen)
Setting
On
Off
On
On
Off
On
4.4 Vol% (corresponds to 100 %LEL)
4.0 Vol% (corresponds to 100 %LEL)
STEL feature – inactive; average dura-
tion = 15 minutes
TWA feature – inactive; average dura-
tion = 8 hours
Acknowledgeable; not latching, pre-
alarm, increasing measured value (for
O
sensor, decreasing measured value
2
as well)
Not acknowledgeable; latching, main
alarm, increasing measured value (for
O
sensor, decreasing measured value
2
as well)
On
Warning
Channel warning
On (180 s)
Device settings
67