4.15

Configuring the device settings

Briefly press
To open the device settings:
1. Select
2. Sign in with the required user level, if necessary.
3. Select and confirm Settings.
4.15.1

Activating day or night mode

1. Open the device settings.
2. Select and confirm Night mode / Day mode.
4.15.2

Changing the device language

1. Open the device settings.
2. Select Changing language....
3. Select and confirm the desired language.
4.15.3

Setting the date and time

1. Open the device settings.
2. Select Date & Time.
3. Select Set date format and then select and confirm the date format.
4. Select Set date and then set and confirm the date.
5. Select Set time and then set and confirm the time.
The user must manually switch between summer and winter time.
When using the X-dock maintenance station, automatic time synchronisation is
possible.
4.15.4

Activating silent mode

Silent mode can be activated for 15 minutes on the gas detector. When silent mode
is active, vibration and the horn are deactivated. Permanent deactivation is possible
using the Dräger CC-Vision PC software.
Silent mode may not be permanently activated for applications according to ATEX.
4.15.5

Activating or deactivating the capture range

1. Open the device settings.
2. Select Capture ranges.
3. Activate or deactivate the capture range.
Dräger recommends that you activate the capture ranges function.
|
Technical manual
X-am® 3500/8000
and
simultaneously to navigate to Monitoring.
in measuring mode and confirm the dialog.