Configuring The Gas Detector; Configuring The Gas Detector With The Dräger X-Dock And Reading The Data Memory - Dräger HFG 000 Series Technical Manual

Table of Contents

Advertisement

Substance
CAS no.
Methyl tert-butyl
1634-04-
ether (MTBE)
4
n-Nonane
111-84-2 Nona
n-octane
111-65-9 Octa
Alpha-pinene
2437-95-
8
Styrene
100-42-5 Styr
Toluene
108-88-3 Tolu
Trichloroethylene
79-01-6
Vinyl chloride
75-01-4
o-xylene
95-47-6
Diesel
68476-
34-6
Petrol
8006-61-
9
Jet fuel
8008-20-
6
1.3-butadiene
106-99-0 BTD1
1-butanol
71-36-3
Isooctane
540-84-1 iOct
1.1-dichloroethene
75-35-4
Tetrahydrofuran
109-99-9 THF
Dimethylformamide 68-12-2
Cyclohexanone
108-94-1 CyHo
PID
-
1) RF: Response factor at 20 °C relative to isobutylene (CAS no.: 115-11-7).
2) The measuring capability of the sensor type is not sufficient for this substance.
7.2

Configuring the gas detector

When the capture range is deactivated, the data memory fills up with log files
significantly faster, making is necessary to read it out more frequently.
When a configuration is transmitted to the gas detector with the CC-Vision PC
software, existing TWA and STEL evaluations are reset.
7.2.1
Configuring the gas detector with the Dräger X-dock and reading
the data memory
To configure a gas detector with the X-dock maintenance station and read the data
memory, insert the gas detector in the X-am 8000 module of the maintenance
station.
Refer to the Data dongle X-dock Instruction for Use for information on how to
configure the gas detector or read the data memory with the maintenance stations.
|
Technical manual
X-am® 3500/8000
PID HC
A1 [ppm] A2 [ppm]
1)
RF
MTBE
0.87
50
1.55
50
1.93
50
aPin
0.38
50
0.44
50
0.56
50
TCE
0.53
50
VC
1.97
50
Xyol
0.55
50
Desl
1
50
Gaso
1.21
50
JetF
1
50
0.69
50
BuOH
4.93
50
1.19
50
DCE
0.85
50
2
50
DMF
0.95
50
0.97
50
PID
1
50
PID LC ppb
A1 [ppm] A2 [ppm]
1)
RF
100
1.06
4
100
2.1
10
100
2.14
10
100
0.48
2
100
0.82
4
100
0.98
4
100
0.95
4
100
2.18
10
100
0.79
4
100
1
4
100
1.21
4
100
1
4
100
0.69
2
100
5.27
10
100
1.45
4
100
0.76
4
100
1.77
10
100
2)
2)
---
---
100
1.08
4
100
1
4
Configuration
8
20
20
4
8
8
8
20
8
8
8
8
4
20
8
8
20
2)
---
8
8
65

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Hfg 001 seriesX-am 3500X-am 8000

Table of Contents