Dräger HFG 000 Series Technical Manual page 27

Table of Contents

Advertisement

2. Mount the pump adapter to the gas detector. Make sure that both guide pins are
in the correct grooves.
mounted correctly, the leak test starts automatically. If the leak test does not
start, the gas detector is not ready for use. Avoid bending the pump adapter.
The gas detector automatically switches to pump mode as soon as the pump
adapter is mounted.
 The leak test starts automatically.
3. When the leak test is displayed, the suction inlet on the probe or hose closes
within 60 s and remains closed until the leak test is complete.
4. Open the suction inlet.
– Leak test successful: Measuring starts. Observe the purging times!
– Leak test failed: Inspect the probe, hose and adapter and repeat the leak
5. Place the probe or the end of the hose on the sampling location.
The temperature at the measurement location may deviate from the temperature
in the gas detector, which may influence the measured value display. The correct
functioning of the temperature correction can only be guaranteed on the gas
detector.
To stop measuring with the pump:
1. Loosen the screw on the pump adapter.
|
Technical manual
X-am® 3500/8000
Check that the pump adapter is mounted correctly. If the pump adapter is
test.
Operation
!
27