Cal sensor

Calibrate the main sensor (adjust zero and sensitivity).
How to access
See Figure 31.
Tree structure
Figure 32: the Calibration menu. Press ESC repeatedly to return to the reading
display.
Menu
Description
Calib. Gas
Configure the value of the calibration gas to be used.
Adjust.'0''
Place the injection hood over the device and inject clean air from
the bottle (flow rate of 30-60 l/h).
Wait for the reading to stabilize (at least 2 minutes).
Press Enter to confirm the zero.
Note: a CO2 sensor pack must always be zeroed using
reconstituted air or nitrogen. Never use ambient air as the zero
since it naturally contains 300-500 ppm of CO
Adjust. 'S''
Place the calibration hood over the detector head and open the
valve on the bottle of calibration gas (flow rate of 30-60 l/h).
The reading displayed will fluctuate until it reaches the stabilization
point. Wait for the reading to stabilize (at least 2 minutes).
Press Enter to confirm the reading.
.