Oldham OLCT 80 User Manual page 49

Gas transmitter
Hide thumbs Also See for OLCT 80:
Table of Contents

Advertisement

Menu
Value displayed
Entry V
Internal value of the main sensor's signal.
Signal V
Main signal in volts.
Meas CH1 V
Signal from the ANA1 sensor in volts.
Meas CH2 V
Signal from the ANA2 sensor in volts.
Temp I °C
Internal temperature within the housing.
Temp C °C
Temperature of the main sensor.
Supply V
Supply voltage to the terminals of the OLCT 80.
Ref V
Internal reference voltage (normally 2.5 V).
1V25 V
Internal reference voltage (normally 1.25 V).
User rate %
Wear rate of the main sensor. A value of 50% represents a 50%
loss of sensitivity. The sensor must be replaced once a 75% wear
rate is reached. This value is recalculated after each calibration.
Output mA
Output current value at the OUT pin (see Figure 5, 2).
3 – Menus
43