Oldham OLCT10N Manual

Digital transmitter

Advertisement

Quick Links

x
P/N : NPO10NGB
Rev A
Digital
Transmitter
OLCT10N

Advertisement

Table of Contents
loading

Summary of Contents for Oldham OLCT10N

  • Page 1 Digital Transmitter OLCT10N P/N : NPO10NGB Rev A...
  • Page 2 Copyright  2010 by Industrial Scientific – Oldham S.A.S First edition, English version All rights reserved. No reproduction of all or part of this document, in any form, is permitted without the written consent of Industrial Scientific – Oldham S.A.S.
  • Page 3: Table Of Contents

    Contents Chapter 1 | Presentation ............6 Composition of the detector ...............6 Type of detector available................7 Chapter 2 | Installation............8 Regulations and conditions of use .............8 Location of the detector................8 Detector positioning ...................8 Connection ....................9 Configuration of communication ..............10 Scope of use ....................12 Chapter 3 | Maintenance ............
  • Page 4 This information shall not, either in whole or in part, by physical, electronic, or any other means whatsoever, be reproduced, copied, divulged, translated, or used as the basis for the manufacture or sale of INDUSTRIAL SCIENTIFIC OLDHAM equipment, or for any other reason without the prior consent of INDUSTRIAL SCIENTIFIC OLDHAM. OLCT10N...
  • Page 5 Warning This is not a contractual document. In the best interest of its customers and with the aim of improving performance, INDUSTRIAL SCIENTIFIC OLDHAM reserves the right to alter the technical features of its equipment without prior notice. READ THESE INSTRUCTIONS CAREFULLY BEFORE THE FIRST USAGE: these instructions should be read by all persons who have or will have responsibility for the use, maintenance, or repair of the instrument.
  • Page 6: Chapter 1 | Presentation

    Chapter 1 | Presentation Purpose The OLCT10N is a digital detector designed to detect a particular gas depending on the type of sensor used. Composition of the detector An OLCT10N detector comprises the following elements: Rep. Description Enclosure Cable gland...
  • Page 7: Type Of Detector Available

    Figure 1 : component parts of an OLCT10N. Type of detector available Range Oxygen (O2) 0-30% vol Carbone monoxide (CO) 0-300 ppm 0-1000 ppm Hydrogen (H2S) 0-30 ppm sulfide 0-100 ppm Nitrogen monoxide (NO) 0-100 ppm 0-300 ppm Nitrogen dioxide (NO2)
  • Page 8: Chapter 2 | Installation

    The detector shall be installed with the cable gland pointing downwards. For explosive gas detectors only, any tilt of more than 45° from the vertical will lead to an inaccurate measurement. OLCT10N Manual...
  • Page 9: Connection

    Connection RS485 transmission There are two cable entries on the OLCT10N, one entry is for the input signal and one is for the output signal – this is used for digitally connecting multiple transmitters on a single line to the MX43 Controller.
  • Page 10: Configuration Of Communication

    (1 to 32) in binary mode. In the illustration to the right, the address 9 (10010) has been defined. The Addressing Table below lists the Figure Switches address configuration. possible combinations. OLCT10N Manual...
  • Page 11 Switches Switches (ON = 1 ; OFF = 0) (ON = 1 ; OFF = 0) Table 1 : Addressing table (address depends on switch positions). Remarks: During module replacement, all the configuration switches of a new ■ module must be positioned in the same configuration as those of the the previous module.
  • Page 12: Scope Of Use

    Inspection or calibration is then recommended. Operation under low oxygen levels If an electrochemical detector sensor is used in an atmosphere comprising less than 1% oxygen for over one hour, the measurement may be an underestimate. OLCT10N Manual...
  • Page 13: Chapter 3 | Maintenance

    Frequency of maintenance Gas detectors are safety devices. OLDHAM recommends the regular testing of fixed gas detection installations. This type of test consists of injecting the standard gas into the detector at a sufficient concentration to activate the pre-set alarms. It is to be understood that this test is in no way a replacement for a detector calibration.
  • Page 14: Automatic Calibration

    The OLCT10N can be calibrated either : Manually Or automatically The manual calibration have to be made only after each sensor or detector replacement. A manual calibration has to be followed by an automatic calibration. Please find below the different calibration gas concentrations.
  • Page 15 Magnetic calibration cap ON Cal ON LED detector status Cal OFF Magnetic calibration OFF Start sensor stabilization Power down if OK [ 1 mn ] Blinking if fault Power down if OK Normal Blinking if fault Cal ON Calibration beginning Cal OFF GREEN Réel...
  • Page 16 Calibration LED Place the magnet horizontally The magnetic calibration allows for one-man and non-intrusive calibration. Manual calibration with zero and span potentiometers is possible by opening the OLCT10N Description Support 6128 972 Slug the gas Magnet 6155771 With a 1...
  • Page 17: Sensor Replacement

    Sensor replacement Sensors should be replaced as a part of regular preventative maintenance or following a failed calibration test. After a sensor replacement, both manual and automatic calibrations have to be conducted To change the sensor: 1. Remove the detector cover 2.
  • Page 18: Manual Calibration

    (in mV) corresponds to the amount of reference gas used. Use the following formula to calculate the correct value for the signal. Stop injecting gas (remove the calibration pipe from the sensor). Wait for the voltmeter to “return to zero.” Reference gas value 1600 (mV) Measuring scale OLCT10N Manual...
  • Page 19 MAINTENANCE WIRES: +VCC (red) = + power supply Signal S (yellow) = signal from 0 mV to 1600 mV for zero and sensitivity settings Ref 2,5V (brown) = zero reference for signal reading from 0 mV to 1600 mV GND (black) = electronic circuit ground Note: Blue and green wires will not be used.
  • Page 20: Chapter 4 | Part Numbers And Accessories

    Chapter 4 | Part numbers and accessories Part Number Description OLCT10N Digital detector, range 0-30% vol O2 OLCT10N-200 OLCT10N Digital detector, range 0-300 ppm CO OLCT10N-204 OLCT10N Digital detector, range 0-1000 ppm CO OLCT10N-205 OLCT10N Digital detector, range 0-30 ppm H2S...
  • Page 21 Part Number Description Kit including OLCT10N Digital detector 6331163 stainless steel bar and a magnetic calibration cap...
  • Page 22: Chapter 5 | Spare Parts

    OLCT10N Digital detector H2S Sensor 6113331 OLCT10N Digital detector NO Sensor 6113332 OLCT10N Digital detector NO2 Sensor 6314118 OLCT10N Digital detector NH3 Sensor range 0-100 ppm 6314119 OLCT10N Digital detector NH3 Sensor range 0- 1000 ppm 6313982 OLCT10N Digital detector O2 Sensor OLCT10N Manual...
  • Page 23: Chapter 6 | Declaration Of Ec Conformity

    Chapter 6 | Declaration of EC conformity...
  • Page 24 OLCT10N Manual...
  • Page 25: Chapter 7 | Specifications

    Chapter 7 | Specifications Detector Dimensions 118*110*60 mm Ingress protection IP65 Enclosure Inlet and outlet cable 2 cable gland M16, diameter 4 to 8 mm Type of cable Type MPI-22A Power supply 12-30V Consumption Electrochemical sensor : 2.5mA in normal operating @ 24V Catalytic sensor : 50 mA in normal operating@ 24V Status indication during...
  • Page 26: Sensor Specifications

    Line length (use with MX43) Below is a table which gives the maximum distances in meter (in feet) of wiring OLCT10N transmitters to the MX43 based on cable size when connected on a single line of without the need of an external power supply.
  • Page 27 Guaranty Plus Points To respond quickly and efficiently to your consultancy needs or order tracking throughout the world via our customer service department. To respond as rapidly as possible to all questions of a technical nature. Quality To assure you of the best quality of our products and service in conformity with the international standards and directives in force.
  • Page 28 For all After Sales Service in France, contact us by email at servicecenter@oldhamgas.com or by telephone at + 33 (0)3 21 60 80 80. For locations near you, please visit us at indsci.com and click on the Oldham Division. OUR MISSION Preserving human life on, above and below the earth Delivering highest quality, best customer service…...