Marking

Safety relevant parameters. Maximum power dissipated
4 W with class T6 or T85°C
10 W with class T5 or T100°C and T4 or T135°C
OLCT 80 d detector (with local sensor, 'd' type)
OLDHAM SAS
62027, ARRAS France
OLCT80 d
CE0080
INERIS 03ATEX0240X
II 2 G D
Ex db IIC T6...T5* Gb
Ex tb IIIC T85°C...T100°C* Db
T.Amb : -20°C to 60°C
WARNING: Do not open when energized. After de-energizing, delay 2 minutes
before opening. Read user manual (cable glands).
T6...T5 ou T85°C...T100°C according to maximum power dissipated
(*)
OLCT 80 D d detector (with remote sensor, 'd' type)
On the detector:
OLDHAM SAS
62027, ARRAS France
OLCT80 D d
CE0080
INERIS 03ATEX0240X
II 2 G D
Ex db IIC T6...T5* Gb
Ex tb IIIC T85°C...T100°C* Db
T.Amb : -20°C to 60°C
WARNING: Do not open when energized. After de-energizing, delay 2 minutes
before opening. Read user manual (cable glands).
T6...T5 ou T85°C...T100°C according to maximum power dissipated
(*)
On the remote sensor:
OLCT 80 D d
CE0080
INERIS 03ATEX0240X
II 2 G D
Ex db IIC T6 Gb
Ex tb IIIC T85°C Db
T.Amb : -20°C to 70°C
WARNING: Do not open when energized.
14 – Special instructions for use in explosive environments and
(*)
:
105
fonctional safety