RELAIS D
Configure the fault relay.
How to access
See Figure 24.
Tree structure
Figure 27: the Fault Relay menu. Press ESC repeatedly to return to the reading
display.
Menu
Description
■
Internal: the relay is triggered by the internal electronics of the
RD intern
transmitter.
■
External: the relay is triggered by the MX 62 central measuring
controller or an API via the RS485 (Modbus) connection.
■
RD sec. pos.
Positive security: the relay will be powered as long as there is no
alarm (positive security). It will be deactivated in the event of a
fault. This option is recommended.
■
Negative security: the relay will be de-energized as long as there
is no fault. The relay will be activated in the event of a fault.
■
Sûre ??
ENTER: confirm the changes made and return to the Alarm/Relay
Configuration menu.
■
ESC: cancel the changes made and return to the Alarm/Relay
Configuration menu.
3 – Menus
38