2.4 Mounting the Sample System (cont.)
The sample system is normally fastened to a flat metal plate that has four
mounting holes. Upon request, GE can also provide the sample system in an
enclosure.
Complete the following steps to mount the sample system:
Fasten the sample system plate or enclosure to a vertical wall or panel
1.
with a bolt in each of the four corners.
Connect the sample system inlet to the process and the outlet to the
2.
return, using appropriate stainless steel fittings and tubing.
CAUTION!
until the probe has been properly installed (see the
following section).

2.5 Installing the Probe

GE probes are usually installed in a sample system to protect the probe from
any damaging elements in the process. The probe is mounted in a
cylindrical container called the
your sample system.
Standard M2 probes and VeriDri probes are mounted into the sample system
or process line with 3/4-16 straight threads that are sealed with an o-ring.
Other fittings are available for special applications.
CAUTION!
process line, without a sample system, consult GE for proper
installation instructions and precautions.
Refer to Figure 24 on page 18, and complete these steps to install the probe
into the sample cell:
Insert the probe into the sample cell and thread the probe into the
1.
sample cell fitting. Make sure you do not to cross the threads.
Tighten the probe securely.
2.
Moisture Target Series 6 Hygrometer User's Manual
Do not start the process flow through the system
sample cell
If the probe is to be mounted directly in the
Chapter 2. Installation
, which is included as part of
17