Chapter 2. Installation

2.3.2 Adapter Plate Mounting

Some customers may need to retrofit an MTS6 into a cut-out sized for GE's
previous OEM panel-mount hygrometers: MTS5, MTS4 or PanaDry
models. The previous generations were of a larger size requiring a 5.40"
(137.2 mm) W x 2.65" (67.3 mm) H cut-out. The MTS6 is smaller, requiring
a 3.69" (94 mm) W x 1.81" (46 mm) H cut-out. For those customers who
need to retrofit the larger panel cut-out sizes, GE offers an optional adapter
plate. See Appendix A, Outline and Installation Drawings, for the required
panel cutout.
Fit the larger gasket around the adapter plate (see Figure 16).
1.
Figure 16: Installing the Adapter Plate Gasket
12
Moisture Target Series 6 Hygrometer User's Manual