GE Moisture Target Series 6 User Manual page 72

Hygrometer
Table of Contents

Advertisement

Chapter 3. Operation and Programming
3.6.4c
Setting the MH Calibration
3.6.4d
Setting the Dew Point Calibration
Press Cancel
62
To set up the MH calibration, highlight Edit
MH and press Enter
screen appears.
Use the left and right arrow keys to select
each digit to be changed and the up and
down arrow keys to increase or decrease its
value. Press Enter
to keep the previous value), and return to the
Edit MH/DP Cal menu.
To set up the dew point calibration, highlight
Edit DP/°C and press Enter
following screen appears.
Use the left and right arrow keys to select
each digit to be changed and the up and
down arrow keys to increase or decrease its
value. Press Enter
to keep the previous value), and return to the
Edit MH/DP Cal menu.
to return to the Cal Data menu.
Moisture Target Series 6 Hygrometer User's Manual
. The following
to save (or Cancel
. The
to save (or Cancel