Selecting The Recorder Output - GE Moisture Target Series 6 User Manual

Hygrometer
Table of Contents

Advertisement

Chapter 2. Installation

2.2 Selecting the Recorder Output

By default, the recorder is set to the current output.
Note:
The customer will provide their own cable for connecting the
Note:
recorder. Acceptable cables range from 16 to 26AWG.
The MTS6 has one isolated analog recorder output. The recorder output
provides either a current or voltage signal, which is set by switch
main PC board.
Complete these steps to check or reset switch
Never connect line voltage or any other power input to
WARNING!
the recorder output terminals.
Make sure the MTS6 is turned off and unplugged.
1.
The MTS must be isolated or disconnected from all voltage
WARNING!
sources before changing the recorder output.
Remove the screw at the top of the back panel (see Figure 4).
2.
4
Figure 4: Back Panel
Moisture Target Series 6 Hygrometer User's Manual
(see Figure 8 on page 7).
S1
on the
S1

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Mts6

Table of Contents