3.6.3 Setting Autocal

To enter the Autocal Settings menu, you must be using a standard
Note:
M Series probe. If a VeriDri probe is being used, AutoCal is not
necessary and will not be accessible.
Moisture Target Series 6 Hygrometer User's Manual
Chapter 3. Operation and Programming
To change the Autocal settings, from the
Settings Menu select AutoCal and press
Enter
. The following screen appears.
To change the Autocal interval settings,
select Interval and press Enter
similar to the following appears.
Use the left and right arrow keys to select
each digit to be changed and the up and
down arrow keys to increase or decrease its
value. Press Enter
to keep the previous value), and return to the
AutoCal Settings menu.
To accept or reject AutoCal, select Cal Now
and press Enter
following appears.
To accept AutoCal, select Yes. To reject
AutoCal select No. Press Enter
confirm your selection and return to the
AutoCal Settings menu.
. A screen
to save (or Cancel
. A screen similar to the
to
59