Selecting The Analog Output Type - GE CGA 351 User Manual

Zirconia oxygen analyzer
Table of Contents

Advertisement

www.DataSheet4U.com
December 2004
Selecting the Analog
Output Type
www.DataSheet4U.com
4-18
Enter the analog output setup menu, as described on page 4-16, and
proceed as follows:
User Aout
Aout Type
Zero Setpoint
Span Setpoint
POWER
ESC
Note: The
[Test], [Zero Trim]
available in the above list box.
User Aout Aout-Type
0-20mA
4-20mA
0-2V
POWER
ESC
Note: The [0-2V] option is software-selected and therefore there is
no need for hardware modifications to output 0-2 volts DC.
Note: The
[Namur]
option is also available in the above list box.
Note: The Namur NE43 specification requires a 4-20 mA output to
be clamped in specific bands to indicate out-of-range errors.
Note: The 0-2V option is software-selected and therefore there is no
need for hardware modifications to output 0-2 volts DC.
The meter now automatically returns to the "Analog Output" prompt.
[ESC]
Either press
until you exit the User Program or proceed to the
appropriate section to continue setting up the analog output.
Use the [ ] and [ ] keys to
[Aout Type]
select
[ENTER]
.
ENTER
[Span Trim]
and
options are also
Use the [ ] and [ ] keys to
select the desired device type,
[ENTER]
and press
ENTER
, and press
.
Setup and Calibration