52 Patient data management

Automatic discharge

Note
The central station automatically discharges patients based on monitoring inactivity or
events from the ADT system. Patients who were entered through an ADT system are
automatically discharged only after the central station receives a discharge or transfer
message.
After receipt of a discharge or transfer message, patients are automatically discharged 4
hours after the last episodic reading is taken or a monitoring session ends.
Patients that were entered into the system without an ADT feed are automatically
discharged 24 hours after the last episodic reading is taken or a monitoring session ends.
Note

Re-admit a patient

1. Click the Patients icon in the Navigation area.
The Patient list appears.
2. Click the Discharged tab.
The list of discharged patients appears.
3. Click on the patient you want to re-admit.
The Patient Details window appears.
4. Enter the new location for the patient and click Save.
The Confirm Changes dialog box appears.
5. Review the information in the dialog box, and if correct, click Yes to return to the
Main Monitoring screen. If further editing is required, click No to return to the
previous screen.
You cannot discharge a patient who is continuously sending data vital signs
measurements to the central station. These patients appear in the list with
the
icon.
This behavior can be customized, enabled, or disabled. Contact your
system administrator to change or modify your settings.
®