Page 1
CP300 12 Channel ECG Recorder Service handbook...
Page 2
4341 State Street Road, PO Box 220 Skaneateles Falls, NY 13153-0220 www.welchallyn.com Sales and Service information: For information about any Welch Allyn product, please call Welch Allyn Technical Support: 1 800 535 6663 + 1 315 685 4560 Australia + 61 29 638 3000...
Symbols Used on the Device ............10 Terms of warranty .............. 10 Introduction ..........11 Standard Features .............. 11 Options ................11 Main Components of the CP300........12 Operating Philosophy ............12 Default Login Codes............13 Keypad................. 14 External Connections............15 2.7.1...
Page 4
Service handbook CP300 Functional Checks ......... 26 Service interval ..............26 General Unit Integrity Check ..........27 6.2.1 Procedure..................27 6.2.2 Continuous Printout (Internal thermal printer only) ......28 Functional Checks and Tests ..........29 6.3.1 External Sight Control ..............29 6.3.2...
Intended Use The CP300 is a 12-channel, ECG device used for the recording, analysis and evaluation of ECG Recordings. Recordings made with the CP300 can be used as a diagnostic aid for heart function and heart conditions. The CP300 is designed for indoor use and can be used for all patients of both sexes, all races, and all ages.
– Ruptured fuses must only be replaced with the same type and rating as the original. Operation with other Devices Use only accessories and other parts recommended or supplied by Welch Allyn. Use of other than recommended or supplied parts may result in injury, inaccurate information and/or damage to the unit.
Safety Notes Service handbook CP300 Safety Symbols and Pictograms Safety Symbols and Pictograms 1.7.1 Symbols Used in this Document The safety level is classified according ANSI Z535.4. The following overview shows the used safety symbols and pictograms used in this manual.
Attention: Consult accompanying documents. Terms of warranty The Welch Allyn CP300 is warranted against defects in material and manufacture for the duration of one year (as from date of purchase). Excluded from this guarantee is damage caused by an accident or as a result of improper handling. The warranty entitles free replacement of the defective part.
Service handbook CP300 Standard Features 2 Introduction The Welch Allyn CP300 is a diagnostic workstation designed to record, display, archive, present, and analyse ECG recordings and other measurements. 12-lead resting ECG recordings give measurements, interpretation, average cycles and rhythm sections as standard. Exercise testing includes predefined stress protocols and dedicated keys for treadmill control.
Operating Philosophy The operating philosophy of the CP300 is that users are allocated user rights which allow access to specific functions. Several user levels are available. It is the system administrator / Manufacturers technician who defines the users and allocates the user level.
Introduction Service handbook CP300 Default Login Codes Default Login Codes There are default login codes for Manufacturers technician and Administrator login levels. Only Manufacturers technician login level has access to unit firmware update, software option update and other menu options. The default logins are as...
Introduction Keypad Keypad 11 12 13 14 The keyboard can be slightly different depending of what version was delivered. Part. nr. 4.150076 = WA PN 714279 Part. No:4.150231 (1) Alpha-numeric keyboard. The keys F1 to F10 have varying functions depending on use (see page 18).
Service handbook CP300 External Connections External Connections All externally connected hardware must be approved by Welch Allyn. Connection of any hardware not approved by Welch Allyn is at the owner‘s risk. The unit guarantee may also be invalid. 2.7.1 Back Panel (1) Four USB connectors (Universal Serial Bus (version 1.1 protocol)).
Introduction Service handbook CP300 The Display The Display Example of the Patient Screen Example of an aquisition Screen (exercise) The display will vary according to the current task being carried out. In all screens however, the top, middle and bottom areas display the same information groups. The following gives examples of the patient screen and the exercise acquisition screen.
• Do not allow the unit to come into contact with acidic vapors or liquids. • The CP300 should not be placed in the vicinity of X-ray or diathermy units, large transformers or electric motors.
Operation Start-up and Initial Preparation 4.1.4 Switching ON and OFF The unit is switched on and off with the On / Off key. 4.1.5 Power Supply and Backup Battery Operation The unit is operated from the mains supply. When mains is connected the mains On / Off Key indicator is lit.
4.2.1 Connecting a Mouse or Trackball Device The CP300 can work with an external point and control device. When a mouse or trackball etc. is connected (to the PS-2 connector on the back panel), the glide point controller operation is disabled.
• Certain glues can react with the paper - do not attach the printout onto a mounting sheet with glue. Welch Allyn can only guarantee perfect printouts when CP300 original chart paper or chart paper of the same quality is used.
The ECG board provides isolated patient connection and initial processing of the ECG signals. The operating system in the CP300 is Windows® NT or XP. The user interface, all settings, graphical presentations etc. are based on Windows®. The evaluation of the ECG data, control of the printer and other functions are realised on the Welch Allyn ECG processor card with its own 68331 microprocessor system.
Page 25
Physical and Functional Overview Service handbook CP300 Functional Overview Note: The connector designations given here are liable to change. 1 2 1 2 1 2 1 2 1 2 3 PC Single Board 1 2 3 1 2 3 Computer...
Every 6 months • General unit integrity check (see following). • The visual, general, measuring and calibration tests and checks ac- By Welch Allyn authorized Every 12 months cording to the checklist at the end of this book (see page 81).
It is not a full functional test but is intended to provide a general confidence check in all the main functional areas. Comprehensive instructions for operating the unit are provided in the CP300 User Guide. This are available from Welch Allyn on request.
Functional Checks General Unit Integrity Check 6.2.2 Continuous Printout (Internal thermal printer only) Manual real-time printout is not available on an external printer because the data processing of inkjet and laser printers is too slow for real time print. When a continuous real-time printout of the ECG is required, it is always printed on the internal thermal printer.
Functional Checks Service handbook CP300 Functional Checks and Tests Functional Checks and Tests 6.3.1 External Sight Control Required equipment • None 6-monthly Check the following: • Mechanical condition of the device: – no cracks or chips in the casing. – mains, patient and all other cable assemblies are in good condition with no crushing, chafing or cuts, etc.).
Functional Checks Functional Checks and Tests 6.3.2 Mains and Battery Indicators (LED) test The internal battery provides power for a limited time. The length of time will depend on many factors enough including battery age, condition, charge, and temperature The battery indicator is lit when running on battery power and blinks when the battery capacity is low.
Functional Checks Service handbook CP300 Functional Checks and Tests 6.3.5 LCD Screen Test • Visually check the screen for spots, or black fields. If many are apparent, the LCD must be replaced (a few faulty pixels is normal). • Check that the LCD shade (contrast and brilliance) is even and the same all over.
3. The paper must stop exactly at the perforation. If this is not the case: – Check that CP300 paper is used. – Clean the paper detection opto window with an alcohol solution.
Functional Checks Service handbook CP300 Functional Checks and Tests 6.3.7 ECG Amplifier and Patient Cable Test (Electrode/ Lead Re- sistance) This gives electrode dc offset and is the voltage drop in the patient cable and electrodes. The result column gives the detected voltage for each electrode in millivolts measured between the electrode on the left leg and each of the individual electrodes.
Functional Checks Functional Checks and Tests 6.3.8 ECG Printout Reference Required Equipment • Calibrated test ECG Patient Simulator (e.g Müller & Sebastiani MS410 ECG Sim- ulator) IMPORTANT! • The measurement devices listed above are subject to the instructions according to ISO Standards in regards to Test Equipment Control.
Page 35
Functional Checks Service handbook CP300 Functional Checks and Tests Reference Printout (Waveform and Intervals) Page 35...
• RS-232 test plug. I/O Test Program The test program is found on the hard disk of the unit: C: Drive (CP300) > Service > Test > Commport.exe CP300 RS Test Plug Using an RS plug, solder together pins 2 and 3.
Failed to open without the unit being plugged in Com 4 will show always “Failed to open”, because it is not used. If “Failed to open” is displayed, the port is occupied by an application running on the CP300. Exit this application and try again. Page 38...
Functional Checks Service handbook CP300 Safety tests Safety tests Required equipment • Safety Tester IEC/EN 60601-1 • Bender Safety Tester (recommended) • HA2000D High voltage measuring unit (recommended) IMPORTANT! The measurement devices listed above are subject to the instructions according to ISO Standards in regards to Test Equipment Control.
QRS Trigger / DC Input / DC Output Function Freeze On Show Start Recording Dialogue Emergency ECG R ti The CP300 does not have a CDROM drive so uploads / downloads the CP300 is over a network, USB memory stick, etc. Page 40...
Software, Firmware and Test Screen Service handbook CP300 Unit Settings Unit Settings 7.1.1 Importing / Exporting / Reset to Default It is possible to export and import the unit settings (color, default speed etc., but not language and some other settings). These can then be used to define the settings of any unit, for example, to set the same unit settings for all units in a department / hospital, etc.
Unzip and copy the file on the desktop / memory stick / network folder. Click the Start Firmware Update icon. The following is displayed: CP300 CP300 Click the browse button to locate the firmware file (.abs file) and install in the normal way.
Software, Firmware and Test Screen Service handbook CP300 Updating the Unit Options Updating the Unit Options The Exercise and Interpretation options are already prepared in the software and are activated by entering an upgrade code. 1. Click the Start Option Update (a) icon. The following screen is displayed:...
When the test screen is entered a code can be entered to perform certain functions. The test screen is a development tool and is not intended for service personnel and the codes given here are for information only. CP300 CP300 The codes are as follows:...
Replacing Major Components Service handbook CP300 Safety Notes 8 Replacing Major Components Safety Notes Danger of electrical shock. Remove mains cable. When working on a open device connect the device via an isolation transformer. Follow the procedures for the prevention of accidents and environmental protection according your national guidelines.
Replacing Major Components Opening the Unit Opening the Unit The top and bottom assemblies are secured with seven recessed screws. Access to the screws is gained from the underside of the unit. To separate, proceed as follows: Turn the unit up-side-down and rest on a soft antistatic cloth. Unscrew and remove the countersunk retaining screws and washers situated in the extreme corners and edges of the unit.
Replacing Major Components Service handbook CP300 Main Board MK 17-1 & PC1-1 Board Main Board MK 17-1 & PC1-1 Board 1. Remove the EMC shield by unscrewing the securing screws. 2. The ETX base board processor os positioned on the PC1-1 single processor board that is located on spacers on the main board (MK17-1).
Replacing Major Components ECG Board MK 16-2 ECG Board MK 16-2 Unscrew the EMC shield and remove Remove the connector to the Main Board (P30) Unscrew the four board retaining screws. Remove the two screws securing the patient connector to the casing. Remove the Board.
Replacing Major Components Service handbook CP300 Reassembling the Unit Reassembling the Unit 8.8.1 Internal Sight Control If the device has been opened, the device must be given a full sight control before it is screwed back together. Check following items: •...
9 Unit Settings Settings at Administrator login level Detailed in this section are all settings available at Administrator login level. 9.1.1 System Settings CP300 System Login/Startup, Users, Institutes, Departments General Filter QRS Trigger / DC Input / DC Output Function Freeze On...
Page 51
Unit Settings Service handbook CP300 Settings at Administrator login level Parameter Options Description Language / Language Select language of choice. This changes the program language immediately Formats without rebooting. All operating system dialogues however, (e.g. printer setup), will remain in the language of the operating system.
Page 52
Select the Remote active box (1) to store recordings remotely (i.e. not on the CP300 database, for example on SEMA-200). select the Local / Wireless box if the SEMA-200 database is installed on the CP300 hard drive. If the SEMA program is installed on a network, do not select this box.
Page 53
Options Description General Check the shut down the computer after quitting the CP300 box to shut down (1) Switch Off Mode the unit directly when the program is exited. If this box is not checked, the unit will enter the Windows environment when the CP300 program is exited. To display a prompt screen before exiting the program, check the Confirm CP300 shutdown box.
If you wish to skip the login screen and go directly into the program when the Login unit is first switched on, check the 'skip the login...' box, and specify the user. When the unit is next switched on the CP300 will automatically login with the defined user. Institutes /...
Unit Settings Service handbook CP300 Settings at Administrator login level 9.1.3 Settings System Login/Startup, Users, Institutes, Departments General Filter QRS Trigger / DC Input / DC Output Function Freeze On Show Start Recording Dialogue Emergency ECG Resting ECG Parameter Options...
Unit Settings Settings at Physician/Administrator level Settings at Physician/Administrator level 9.2.1 General Filter Settings General Filter QRS Trigger / DC Input / DC Output Function Freeze On Show Start Recording Dialogue Emergency ECG Resting ECG Exercise ECG ECG View Myogram On / Off - The Myogram filter reduces muscle induced noise.
Unit Settings Service handbook CP300 Settings at Physician/Administrator level 9.2.2 QRS /Trigger / DC Input / DC Output Settings General Filter QRS Trigger / DC Input / DC Output Function Freeze On Show Start Recording Dialogue Emergency ECG Resting ECG...
Unit Settings Settings at Physician/Administrator level 9.2.3 Function Freeze The freeze function is an easy way of saving screen settings so that when the screen Settings is next visited, the same display settings (e.g. lead order, print settings, etc.,) are set. General Filter When Freeze is enabled (select before the option), an extra freeze icon appears in the QRS Trigger / DC Input / DC Output...
Unit Settings Service handbook CP300 Settings at Physician/Administrator level 9.2.5 ECG Settings Filter and Lead Order Note that all of these settings are also available in the settings menu of the recording screens. Settings General Filter QRS Trigger / DC Input / DC Output...
Unit Settings Settings at Physician/Administrator level 9.2.6 ECG Settings Printer and Printout settings A printout can be obtained on the internal, or an external printer, directly after an auto Settings recording has been made. General Filter QRS Trigger / DC Input / DC Output Function Freeze On The format of the printout for both the internal and external printer is defined by the Show Start Recording Dialogue...
Unit Settings Service handbook CP300 Settings at Physician/Administrator level 9.2.7 Exercise ECG Settings Note that all of these settings are also available in the settings menu of the recording screens. Settings General Filter QRS Trigger / DC Input / DC Output...
Exercise ECG ECG View Select the setting of the load to maximum, or interpolate the load to that of the last and second to last load stages. An explanation of interpolation is given in the CP300 user guide. Page 62...
Unit Settings Service handbook CP300 Settings at Physician/Administrator level 9.2.9 Defining a Protocol Select the protocol option from the Exercise ECG menu. All of the defined protocols are listed. To define a new protocol click the Add icon (1). To view a previously defined protocol, highlight it in the list and click the Settings icon (2).
Page 64
Unit Settings Settings at Physician/Administrator level Exercise Phase Ramp - A ramp protocol means that the load increase is applied gradually over the duration of the test at a rate of x Watts per minute. When this option is selected, the start load and the load increase per minute must be entered.
The casing of the CP300 can be cleaned with a soft damp cloth on the surface only. Where necessary a domestic non-caustic cleaner can be used for grease and finger marks.
Ensure that fresh CP300 paper is installed. • Old paper inserted. not clear, or the Note that the thermal paper used for the CP300 is heat and light printout is sensitive. If it is not stored in its original seal, stored in high temper- ‘patchy‘.
Maintenance Service handbook CP300 Master Reset 10.3 Fault Possible Causes and indicators Remedies and Fault Location Patient recordings Connection to network lost, check network. • No connection with SEMA data- shown base In system settings Settings > system > database (administrator...
Maintenance 10.4 Changing the Mains Fuse 10.4 Changing the Mains Fuse Before the fuse and mains voltage are changed, the device must be disconnected from the mains and the mains plug removed for the wall socket. The fuse may only be replaced by the fuse type the table below. 10.4.1 Fuse Types Voltage range...
Technical Data Service handbook CP300 System 11.1 11 Technical Data 11.1 System Dimensions 100 x 285 x 350 mm, (0.328 x 0.935 x 1.148 feet) 6.8 kg (14.99 lbs) Monitor • Backlit for graphic and LCD alphanumeric representation • 15" TFT- display, high-resolution 1024 x 768 pixels...
Technical Data 11.4 EMC Information 11.4 EMC Information The unit meets the Collateral Standards of Electromagnetic compatibility – Require- ments and tests IEC/EN 60601-1-2 the limits and methods of measurement of elec- tromagnetic disturbance characteristics of industrial, scientific and medical radio fre- quency equipment.
Technical Data Service handbook CP300 EMC Information 11.4 11.4.3 Emissions Equipment and Systems - Table 204 Emissions Equipment and Systems that are NOT Life-Supporting Immunity Test IEC 606101 Test Level Compliance Level Electromagnetic environment guidance Portable and mobile communications equipment...
Technical Data 11.4 EMC Information 11.4.4 Recommended Separation Distances - Table 206 Recommended Separation Distance between Portable and Mo- bile RF Communications Equipment and the Device The user of the device can help prevent electromagnetic interference by maintaining a mini- mum distance between portable and mobile RF communications equipment (transmitter) and this device as recommended below, according to the maximum output power of the communi- cation equipment...
Construction Drawings Service handbook CP300 Upper Assembly 12.1 12 Construction Drawings Conversion from drawing number to Welch Allyn PN refer to DIR30034004. 12.1 Upper Assembly Isolation Shield Deckel 4.415 542 16 x ZK M2.5x4 2 x PT22x8 4.910 200 4.910 227...
CP300 checklist table 13.1 13 Checklist The following procedural checklist must be carried out by authorized Welch Allyn trained personnel. The recommended interval is that the unit is checked every 12 months. It is a requirement of EN 60601 that the unit undergoes a safety test at defined intervals (see page 37), and a complete functional check at least every 24 months (detailed in the following checklist).
Page 82
Checklist 13.1 CP300 checklist table Reference False Remarks 6.3.5 LCD Screen Test (page 31) No spots or black fields on the screen. LCD shade (contrast and brilliance) is even all over. 6.3.6 Printer Checks (page 32) No fading Alignment OK...
Page 83
Checklist Service handbook CP300 CP300 checklist table 13.1 Reference False Remarks 6.4 I / O Port Checks (page 37) 1. Com 2 port I/O ok - both green lights lit 2. Com 4 port I/O ok - both green lights lit Ext 1, 2 and 3 ok - both green lights lit for all ports 6.5 Safety tests (page 39)