Change Set Up Menu

To access Change Set Up Menu:
1. Turn the pump on in the normal operation mode.
2. To access the Change Set Up Menu, press the INFO key multiple times until
the "Change Set Up" option is displayed,(it is always displayed as the 4th
option).
3. Select "Change Set Up", press OK.
4. Scroll down to the option "More", press OK.
5. Enter access "Change Set Up" Code, press OK.
The Change Set Up Menu is intended for use by clinical personnel and authorized
service technicians.
Item Parameter
1
restart pump
2
pressure default
3
select program
4
priming rate
5
max bolus volume
6
titration bolus
7
bolus rate
8
maximal rate
9
KVO rate
10
flow control
11
set loading test
12
air bubble size
13
store last program
14
operation LED
15
time option
16
delay before start
17
language
18
set time and date
19
history
Description
Allows clinical personnel to return pump to normal operation mode.
Allows clinical personnel to set the default occlusion alarm pressure.
The pump will default to this pressure setting each time the pump is
started.
Allows clinical personnel to select program from list (Continuous,
Intermitent, PCA, TPN, or 25 steps).
Allows clinical personnel to set priming rate.
Allows clinical personnel to set maximum bolus volume (0-100 ml).
Allows clinical personnel to set titration bolus (On/ Off).
Allows clinical personnel to set bolus rate (0-1200 ml/h).
Allows clinical personnel to set maximal rate (0.1-1200 ml/h).
Allows clinical personnel to set KVO rate (0.0-5.0 ml/h).
Allows clinical personnel to set flow control (0 - 20 drops/ml).
Allows clinical personnel to activate set loading test (On/Off).
Allows clinical personnel to set air bubble size (0.1 - 2.0 ml).
Allows clinical personnel to store last program (On/Off).
Allows clinical personnel to set operational indicator LED (On/Off)
Allows clinical personnel to set time option (On/Off).
Allows clinical personnel to set delay before start (On/Off).
Allows clinical personnel to choose language (English, Spanish).
Allows clinical personnel to date and time (dd, mm, yy, hh, mm).
Allows clinical personnel to review event history settings and alarms
(for last 500 events)