If the screen displays "Failure", repeat step
After Arc correction and splice position correction complete, press "
to exit Arc correction mode.
ATTENTION:
Cleave Angle Limit is solely set under the Arc correction mode, which is
►
irrelevant with the one under splice mode.
Arc correction often needs to be operated a couple of times, user should
►
follow the steps patiently.
3.3
Cleaning of Fibre Fusion Splicer
3.3.1
Clean the V-groove
The presence of impurities in V-groove will make the fibre image deviate from
the normal position, resulting in dis-alignment, causing more splice loss. So
users should regularly check and clean V-groove, the specific process is as
follows:
Open the windproof cover of Fibre fusion splicer.
Use the specific brush to clean the impurities in V groove bottom as
shown in vew Figure.
Please use the fine cotton swab dipped in alcohol to clean the bottom of
the V-groove. as shown in Figure 31 (b).
30
Page
a
Figure 31. Clean the V-groove.
.
b