TESTO 6681 Profibus Instruction Manual page 86

Humidity transmitter with profibus module testo 6610. probes p2a software. parameterizing, adjusting and analyzing software
Hide thumbs Also See for 6681 Profibus:
Table of Contents

Advertisement

testo 6681 Profibus
86
Message
00E00 T process high**
02806
Condensation*
02807
Values less than
0 % RH**
02809
Sensor early
warning*
*
Early warning
**
Current malfunction
- 1.5 Status, warning and error messages
Display
The process temperature
exceeds the temperature
designated for the probe
100 % RH has been
reached, condensation
developing
The adjustment or sensor
is faulty
For testo 6617 probe only:
The cover electrode of the
sensor is damaged; this
may soon cause the sensor
to break
Cause
Remedying of fault
Remove the probe from
the process and take any
necessary measures to
lower the process
temperature
Take measures to reduce
process humidity
Check adjustment (via
P2A adjustment history,
perform 2-point
adjustment where
necessary)
If the problem persists,
contact Testo Service
Carry out visual
inspection
If the mirror-like surface of
the sensor is dirty or
damaged, contact Testo
Service