testo 6681 Profibus
82
Press SET, edit value: Scroll one digit to the right using , increase
7
value of digit by 1 using . Confirm with SET or abort entry with ESC.
Continue to Analog Adj. Ch. 1 with or .
8
9
Continue with the adjustment of the analog outputs (see below, step 2) or
press ESC to return to Main Menu Adjustment.
10 Continue to Main Menu Reset using or or return to Measuring
Mode by pressing ESC.
Performing the 2-Points-Adjustment for 20% RH and 80 % RH
In the Measuring Mode press SET, select main menu Adjust with or
1
and confirm selection with SET.
Continue to 2Pt Adjust 20/80% with or .
2
3
Press SET.
Continue to Adj. Point 20% or Adj. Point 80% with or
4
5
Confirm selection with SET.
The red LED ADJ flashes
In the display 2Pt Adjust 20/80%, 1-point adjustment, Probe reset
appears consecutively.
The adjusted measurement values appear after probe reset.
Performing analog adjustment
1
In Measuring Mode, press SET, select Main Menu Adjust using or
and confirm selection with SET.
Select Analog Adj. Ch. 1 using or and confirm with SET.
2
Select Adj. Point 1 using or .
3
4
Press SET. Read off multimeter display (e.g. 5.601 mA) and enter this
value in the user menu. Do this by scrolling one digit to the right using
and increasing the value of digit by 1 using . Confirm with SET or abort
entry with ESC.
- 1.4 Operation
Please refer to chapter 1.3.6.4, page 61.
Each channel is adjusted at three points in the analog range
(at 10 %; 50 %; 90°% of the analog scale).