testo 6681 Profibus
72
1.4.6 The individual main menus
1.4.6.1
Editing main menu of channel 1
An overview is given in Overview of the testo 6681 user menu, chapter 1.4.5,
page 70).
You can perform basic settings for channel 1.
In Measuring Mode, press SET, select Main Menu Channel 1 using or
1
and confirm selection with SET.
One of the following parameters can now be selected using or , after
which the selection must be confirmed with SET:
Channel 1 unit
The parameter for this channel is selected.
Selection: % RH, °C, °F, °Ctd, °Ftd, g/m³, gr/ft³, g/kg, gr/lb, kJ/kg,
BTU/lb, °Ctw, °Ftw, "H2O, hPa, ppm
Edit/select parameter with or , confirm with SET or abort input
with ESC.
Scale minimum for channel 1
The lower scale limit is edited;
Unit as selected above (example: 4 mA = 0 % RH).
Editing the value: Scroll one digit to the right using , increase value
of digit by 1 using . Confirm with SET or abort entry with ESC.
Scale maximum for channel 1
The upper scale limit is edited;
Unit as selected above (example: 20 mA = 100 % RH).
Editing the value: Scroll one digit to the right using , increase value
of digit by 1 using . Confirm with SET or abort entry with ESC.
Signal delay ("Damping") for channel 1
The analog signal can be delayed ("Damping"); a time constant is
selected for this (1 = no delay; 15 = longest delay).
Edit/select parameter using or , confirm with SET or cancel entry
via ESC.
Continue to Main Menu Channel 2 using or or return to Measuring
2
Mode by pressing ESC.
1.4.6.2
Editing main menu of channel 2
See channel 1.
- 1.4 Operation