Warning Messages - TESTO 6681 Profibus Instruction Manual

Humidity transmitter with profibus module testo 6610. probes p2a software. parameterizing, adjusting and analyzing software
Hide thumbs Also See for 6681 Profibus:
Table of Contents

Advertisement

Message
02d07
Probe disconnected
02104
Analog out adjust
02101
1-point adjustment
02102
2-point adjustment
11.3%
02103
2-point adjustment
75.3%
02120
2-point adjustment 20 % As part of the 2-point adjustment, an adjustment
02130
2-point adjustment 80 % As part of the 2-point adjustment, an adjustment
02105
Self-adjustment active
02518
Probe reset

1.5.2 Warning messages

Warning messages show an early warning or a current malfunction which
may negatively impact measuring.
Message
Display
02101
2-point adjustment
drift*
00E00 T ambient high**
00E01 T ambient low**
00E02 Supply voltage
low**
testo 6681 Profibus
Display
No probe is connected
An analog adjustment has been made
A 1-point adjustment is performed.
As part of the 2-point adjustment, an adjustment
is performed at 11.3 % RH
As part of the 2-point adjustment, an adjustment
is performed at 75.3 % RH
is performed at 20 % RH
is performed at 80 % RH
For testo 6615 probe only: The probe performs
an automatic self-adjustment
Probe reset: The probe performs a reset
In the 2-point adjustment,
corrections repeatedly
occur in the same direction;
this may indicate a sensor
drift
The ambient temperature
exceeds the permissible
temperature for the
transmitter
The ambient temperature is
below the permissible
temperature for the
transmitter
The supply voltage is below
the minimum voltage
required for the transmitter
- 1.5 Status, warning and error messages 85
Description
Cause
Remedying of fault
Send the probe into Testo
Service
Take necessary
measures to lower
ambient temperature, e.g.
through venting or cooling
Take necessary
measures to raise
ambient temperature, e.g.
through heating
Ensure sufficient voltage
supply