Chapter 27 – Basal-IQ PLGS Technology Overview
4. Basal insulin delivery is resumed if
insulin delivery has been suspended
for 2 hours in a 2.5 hour window.
For example: If insulin is suspended
for 2 hours, it will resume for at least
30 minutes. If after 30 minutes have
passed, rule 1 or 2 above is true,
insulin will be suspended again for
another 30 minutes.
NOTE: Diagrams are
Representations Only
The diagrams depicted here are sample repre-
sentations only, and not to be interpreted as
actual system performance.
NOTE: Calibrating a Sensor While
Insulin is Suspended
If you need to calibrate your sensor while Basal-
IQ is currently active, and insulin deliveries have
been automatically suspended, insulin delivery
will automatically be resumed if the CGM sensor
value is above 70 mg/dL. The Basal-IQ Technol-
ogy requires three new CGM sensor values to
make a prediction after a sensor calibration.
250

27.3 Turn Basal-IQ On and Off

Once you have a CGM sensor session
started, and the CGM has been cali-
brated, the Basal-IQ Technology can
be turned on or off by following the
steps below.
1. From the Home Screen, tap
OPTIONS.
2. Tap the Down Arrow.
3. Tap Basal-IQ.
4. Tap the ON/OFF toggle.
5. Tap SAVE.
Once turned on, the Basal-IQ Technol-
ogy alert settings are displayed. The
Suspend Alert and the Resume Alert
can be set to on or off depending on
your preference. By default alerts are
set to the off position. See Chapter
Basal-IQ Alerts
to learn more about
these alerts.
NOTE: Turning Basal-IQ Off
During a Suspension
If the Basal-IQ Technology is active, and insulin
delivery has been suspended, and you turn the
feature off, insulin delivery will be resumed at
the current profile rate.