Tandem t:slim X2 User Manual page 275

Insulin pump with basal-iq technology
Hide thumbs Also See for t:slim X2:
Table of Contents

Advertisement

Specification Type
Basal Delivery Accuracy at all Flow
Rates (tested per IEC 60601-2-24)
Bolus Delivery Accuracy at all Volumes
(tested per IEC 60601-2-24)
Patient Protection from Air Infusion
Maximum Infusion Pressure Generated
and Occlusion Alarm Threshold
Frequency of Basal Delivery
Retention Time of Electronic Memory
when Internal System Battery is Fully
Discharged (including Alarm Settings
and Alarm History)
Infusion Set used for Testing
Typical Operating Time when System is
Operating at Intermediate Rate
Customer Technical Support: (877) 801-6901
Specification Details
±5%
The System is designed to vent automatically when there is a pressure difference between inside the cartridge
and the surrounding air. In certain conditions, such as a gradual elevation change of 1,000 feet, the System may
not vent immediately and delivery accuracy can vary up to 15% until 3 units have been delivered or elevation
changes by more than 1,000 feet.
±5%
The pump provides subcutaneous delivery into interstitial tissue and does not deliver intravenous injections.
Clear tubing aids in detecting air.
30 PSI
5 minutes for all Basal Rates
Greater than 30 days
Unomedical Comfort
Infusion Set
During normal use, the intermediate rate is 2 units/hr; battery charge can be reasonably expected to last up to 7
days (5 days if using CGM) from a fully charged state to a totally discharged state
t:slim X2 Pump Specifications (Part 2 of 3)
Chapter 32 – Technical Specifications
273