Chapter 33 – Basal-IQ Technology Clinical Study Overview
Table 11. Insulin Resumption Accuracy
The following table provides the accuracy of the PLGS Algorithm's Resume Actions relative to the corresponding YSI values. The
Resume Actions were analyzed at the first opportunity to resume insulin based on the YSI data, then 5 minutes later and 10 minutes
later.
TRUE Resumes
FALSE Resumes
MISSED Resumes
Total Events
310
Resume Actions (%)
0 minutes
1,356 (51.42%)
1,356 (65.57%)
332 (12.59%)
949 (35.99%)
2,637 (100.00%)
2,068 (100.00%)
+5 minutes
+10 minutes
1,356 (73.06%)
332 (16.05%)
332 (17.89%)
380 (18.38%)
168 (9.05%)
1,856 (100.00%)