Chapter 25 – CGM Alerts and Errors
25.4 Incomplete Calibration
What will I see on the screen?
INCOMPLETE CALIBRATION
This CGM calibration has not
been completed.
CLOSE
222
What does it mean?
If you start to enter a calibration value using the keypad and do not complete the entry
within 90 seconds, this screen appears.
How will the System notify me?
2 beeps or vibrations depending on Pump Volume selected.
Will the System re-notify me?
Yes, every 5 minutes until confirmed.
How should I respond?
Tap CLOSE and complete your calibration by entering the value using the onscreen
keypad.
Customer Technical Support: (877) 801-6901