15.5 Incomplete Load Sequence Alerts

Incomplete Cartridge Change Alert
What will I see on the screen?
CHANGE CARTRIDGE ALERT
The cartridge loading process has
not been completed.
CLOSE
144
What does it mean?
You selected Change Cartridge from the Load menu but did not complete the process
within 3 minutes.
How will the System notify me?
2 sequences of 3 notes or 2 vibrations depending on the volume/vibrate setting
selected in Pump Volume.
Will the System re-notify me?
Yes, every 5 minutes until acknowledged.
How should I respond?
Tap CLOSE. Complete the cartridge change process.
Customer Technical Support: (877) 801-6901