Tandem t:slim X2 User Manual page 137

Insulin pump with basal-iq technology
Hide thumbs Also See for t:slim X2:
Table of Contents

Advertisement

6. Verify Auto-Off is set to on, then tap
Time.
7. Using the onscreen keypad, enter
the number of hours (from 5 to 24
hrs) that you want the Auto-Off
Alarm to be triggered, and tap
DONE.
8. Tap DONE, then tap SAVE when all
changes are complete.
9. Tap Tandem Logo to return to the
Home Screen.
To Respond to Auto-Off Warning
Tap DO NOT SHUT DOWN.
WARNING
Auto-Off alarm will shut pump
down if no action is taken.
No buttons pressed in:
12 hrs
Pump will shut down in:
30 sec
DO NOT SHUT DOWN
Customer Technical Support: (877) 801-6901
Chapter 14 – User Settable Alerts and Alarms
 The warning clears and the pump
returns to normal operation.
If you do not clear the warning within
the 30-second countdown period, the
AUTO-OFF ALARM occurs, accompa-
nied by an audible alarm. This alarm
notifies you that your pump has
stopped delivering insulin.
Auto-Off Alarm Screen
Tap CLOSE.
AUTO-OFF ALARM
ALL DELIVERIES STOPPED!
No activity for more than
6 hours.
CLOSE
 The Home Screen appears,
indicating a status of "All Deliveries
Stopped."
You must resume delivery to continue
therapy, refer to Section
8.2 Resuming
Insulin
Delivery.
135