Advertisement

5. Using the onscreen keypad, enter
the blood glucose value from your
meter.
PRECAUTION
DO enter the exact BG value displayed on
your meter within 5 minutes of using your
meter. Don't enter the Dexcom G6 reading
for calibration.
6. Tap DONE.
7. Tap YES to confirm the calibration.
Tap NO if the BG value does not
exactly match the reading from your
meter. The onscreen keypad will
appear. Enter the exact reading
from your meter.
 The "CALIBRATION ACCEPTED"
screen will appear, followed by the
CGM Home Screen or bolus
prompt.
22.5 Other Reasons You May Need
to Calibrate
You may need to calibrate when your
System did not accept the last calibra-
Customer Technical Support: (877) 801-6901
Chapter 22 – Calibrating Your CGM System
tion, or when the blood glucose value
you entered for calibration is very differ-
ent from the sensor glucose reading.
When you see the CALIBRATE CGM
screen, calibrate using the instructions
in the previous chapters.
If you see the CALIBRATION ERROR
screen, you will be prompted to enter a
blood glucose value to calibrate in
either 15 minutes or 1 hour, depending
on the error.