Max Basal Alert 2
What will I see on the screen?
MAX BASAL ALERT
You have exceeded 2x your highest
basal setting. Please review your
current temp rate in the Options
menu.
CLOSE
Customer Technical Support: (877) 801-6901
What does it mean?
An active temp rate exceeded 2 times your highest basal setting defined in your Per-
sonal Profile.
How will the System notify me?
2 sequences of 3 notes or 2 vibrations depending on the volume/vibrate setting
selected in Pump Volume.
Will the System re-notify me?
Yes, every 5 minutes until acknowledged.
How should I respond?
Tap CLOSE and review your current temp rate in the Options menu.