Flow Sensor Calibration - Badger Meter FC-5000 User Manual

Flow display. industrial flow computer
Hide thumbs Also See for FC-5000:
Table of Contents

Advertisement

Advanced Setup
Option
Description
• Unique to products with raw reed switches
• Pulse input channel
PULSE:
DEBOUNCE
• Single K-Factor entry
K-FACTOR
• Passive sensor: Pullup resistor to 12V for excitation
• Unique to the Data Industrial (DIC) product line
• Pulse input channel
PULSE:
DIC
• K & Offset values entered
• Active sensor: No pullup resistor

Flow Sensor Calibration

The flow sensor calibration parameters vary with the flow sensor type selected�
1� Navigate to SETUP > ADVANCED SETUP > SENSOR INPUTS�
2� Highlight FLOW SENSOO CAL, then press ENTEO�
N
OTEE:
For Dual pulse input versions, highlight FLOW SENSOO 1 CAL or FLOW
SENSOO 2 CAL�
3� Select and edit each option, as necessary� The options include:
Option
Description
K-FACTOR
A singular K-factor entry point�
OFFSET
Used to apply an offset to sensor input calibration
LOW FLOW
The point at which the display reads zero� Represented in configured unit of
CUTOFF
measure
Appears when a "MULTI-POINT CAL" option is picked for a flow sensor type, to
MULTI-POINT
replace a single K-Factor� Allows you to program a
TABLE
linearization table�
10 Point Linearization
The FC-5000 Flow Display can be set up to linearize the output from an eligible
flow meter� The calibration data for a particular flow meter are included when the
meter, calibration and FC-5000 unit are ordered from the factory� The calibration
data is represented as FREQUENCY (f: Hz) vs K-FACTOR (KFct)�
Page 30
Any pulse producing sensor coupled with a reed switch
Example: Industrial Oval Gear
Pulses per unit of volume
Any pulse producing sensor
Examples: TTL, RF carriers w/ amplifier
K = unit of volume per pulse
Table 13: Flow sensor types
Table 14: Flow sensor calibration options
CTL-UM-02765-EN-01
June 2018