User Setup Programming; Flow Units Selection; Tube Size Programming - Badger Meter RCM 2000 Installation And Operation Manual

Research control mag meter low flow magnetic flow meter
Table of Contents

Advertisement

Programming
7.3

User setup programming

7.3.1 Flow units selection

7.3.2 Tube size programming

Installation and operation manual • Version RCM-10/03-e
These menus are entered by first pressing the Enter
key.
Use the up arrow to move the > prompt next to Setup
location on the screen, then press the right arrow key
to select the next set of menus.
The first setup menu selection is the flow unit
selection menu. Use the up arrow to move the >
prompt to the change location on the screen, then
press the right arrow key to select the next menu.
Use the right arrow to select the flow rate units
(GPM, LPM, CCM). When selecting the flow units the
velocity units are automatically selected based upon
whether they are metric (mps) or English (fps). To
save the units press the Enter key. This action will
automatically move to the next menu.
Note. When changing units remember to check the
full scale flow and the pulse output and change to the
desired values. It is also a good practice to reset the
totalizer to prevent mixed units in the totalizer value.
This menu selection changes the flow tube I.D. Use
the up arrow to move the > prompt to the change
location on the screen, then press the right arrow key
to select the next menu.
Use the up or right arrow to adjust the tube I. D. The
tube size is always represented in mm. To save the
value, press the Enter key. This action will
automatically move to the next menu.
Note: If the tubes size is changed the sensor
configuration file must be downloaded through the
serial port to ensure maximum accuracy.