The intended use of the P-725 is only possible in a completely assembled and connected state. The P-725 must be operated with a suitable controller that is available from PI. The controller is not included in the scope of delivery of the P-725.
Unsuitable cables can damage the electronics. Only use cables from PI for connecting the P-725 to the electronics. Physik Instrumente (PI ) GmbH & Co. KG, Auf der Roemerstr. 1, 76228 Karlsruhe/Germany Phone +49 721 4846-0, Fax +49 721 4846-1019, Email info@pi.ws,...
Figure 1: Components of the Z drive, example of a P-725.2CD, here without symbol for protective earth connection Base body Cable exit Protective earth connection Threaded aperture Physik Instrumente (PI ) GmbH & Co. KG, Auf der Roemerstr. 1, 76228 Karlsruhe/Germany Phone +49 721 4846-0, Fax +49 721 4846-1019, Email info@pi.ws, www.pi.ws...
Page 11
3. Affix the P-725 to the microscope using four suitable M2.5 screws and the mounting holes in the base body of the P-725 (see Figure 3 above). Physik Instrumente (PI ) GmbH & Co. KG, Auf der Roemerstr. 1, 76228 Karlsruhe/Germany Phone +49 721 4846-0, Fax +49 721 4846-1019, Email info@pi.ws,...
Screw the objective into the P-725. See the P721T0002 technical note (p. 7) for details. Pay attention to a maximum torque of 0.6 Nm. Physik Instrumente (PI ) GmbH & Co. KG, Auf der Roemerstr. 1, 76228 Karlsruhe/Germany Phone +49 721 4846-0, Fax +49 721 4846-1019, Email info@pi.ws,...
The P-725 is operated near its resonant frequency. If you notice oscillations, stop the P-725 immediately. Physik Instrumente (PI ) GmbH & Co. KG, Auf der Roemerstr. 1, 76228 Karlsruhe/Germany Phone +49 721 4846-0, Fax +49 721 4846-1019, Email info@pi.ws,...
Discharging a P-725 that is not connected to the controller Connect the P-725 to the switched-off PI controller for 10 seconds. Physik Instrumente (PI ) GmbH & Co. KG, Auf der Roemerstr. 1, 76228 Karlsruhe/Germany Phone +49 721 4846-0, Fax +49 721 4846-1019, Email info@pi.ws,...
(e.g., ethanol or isopropyl alcohol). Do not do any ultrasonic cleaning. Customer Service For inquiries and orders, contact your PI sales engineer or send us an email (info@pi.ws). If you have questions concerning your system, have the following information ready: −...
Unloaded resonant as P-725.xCL ±20 % frequency Resonant frequency as P-725.xCL ±20 % at 150 g Physik Instrumente (PI ) GmbH & Co. KG, Auf der Roemerstr. 1, 76228 Karlsruhe/Germany Phone +49 721 4846-0, Fax +49 721 4846-1019, Email info@pi.ws, www.pi.ws...
Page 17
*Specifications assured from 17° C to 23° C. The piezo scanner has to be recalibrated for temperatures outside this range. Physik Instrumente (PI ) GmbH & Co. KG, Auf der Roemerstr. 1, 76228 Karlsruhe/Germany Phone +49 721 4846-0, Fax +49 721 4846-1019, Email info@pi.ws,...
*Specifications assured from 17 °C to 23 °C. The piezo scanner has to be recalibrated for temperatures outside this range. Physik Instrumente (PI ) GmbH & Co. KG, Auf der Roemerstr. 1, 76228 Karlsruhe/Germany Phone +49 721 4846-0, Fax +49 721 4846-1019, Email info@pi.ws,...
Dimensions in mm. Note that the decimal places are separated by a comma in the drawings. P-725.xCD, P-725.xCL, P-725.x0L Figure 6: P-725.xCD, P-725.xCL, P-725.x0L (P-721.xxQ adapter to be ordered separately) Physik Instrumente (PI ) GmbH & Co. KG, Auf der Roemerstr. 1, 76228 Karlsruhe/Germany Phone +49 721 4846-0, Fax +49 721 4846-1019, Email info@pi.ws, www.pi.ws...
Cable shield part of the capacitive sensor) Piezo voltage Ground of piezo voltage on cable shield Physik Instrumente (PI ) GmbH & Co. KG, Auf der Roemerstr. 1, 76228 Karlsruhe/Germany Phone +49 721 4846-0, Fax +49 721 4846-1019, Email info@pi.ws, www.pi.ws...
Dispose of your old equipment according to international, national, and local rules and regulations. In order to fulfil its responsibility as the product manufacturer, Physik Instrumente (PI) GmbH & Co. KG undertakes environmentally correct disposal of all old PI equipment made available on the market after 13 August 2005 without charge.