Measures for avoiding the risk. INFORMATION Additional information that can affect your application. Physik Instrumente (PI ) GmbH & Co. KG, Auf der Roemerstrasse 1, 76228 Karlsruhe, Germany Page 4 / 40 Phone +49 721 4846-0, Fax +49 721 4846-1019, Email info@pi.de,...
5. Click the desired manual and fill out the inquiry form. The download link will then be sent to the email address entered. Physik Instrumente (PI ) GmbH & Co. KG, Auf der Roemerstrasse 1, 76228 Karlsruhe, Germany Page 5 / 40 Phone +49 721 4846-0, Fax +49 721 4846-1019, Email info@pi.de,...
It is only possible to use the P-725 as intended when it is completely mounted and connected. It must be operated with a suitable controller that is available from PI (p. 10). The controller is not included in the scope of delivery of the P-725.
Page 7
Do not use extension cables. If you need longer cables, contact our customer service department (p. 30). Physik Instrumente (PI ) GmbH & Co. KG, Auf der Roemerstrasse 1, 76228 Karlsruhe, Germany Page 7 / 40 Phone +49 721 4846-0, Fax +49 721 4846-1019, Email info@pi.de,...
Page 17
10. Screw the objective from below into the thread adapter on the objective side of the PIFOC positioner. Do not exceed the maximum torque of 0.6 Nm. Physik Instrumente (PI ) GmbH & Co. KG, Auf der Roemerstrasse 1, 76228 Karlsruhe, Germany Page 17 / 40 Phone +49 721 4846-0, Fax +49 721 4846-1019, Email info@pi.de,...
M26 × 1/36” 23 mm P-721.11A M25 × 0.75 mm 22 mm Physik Instrumente (PI ) GmbH & Co. KG, Auf der Roemerstrasse 1, 76228 Karlsruhe, Germany Page 18 / 40 Phone +49 721 4846-0, Fax +49 721 4846-1019, Email info@pi.de, www.pi.ws...
Page 19
Use the key tool supplied for tightening the thread adapter on the objective side. Physik Instrumente (PI ) GmbH & Co. KG, Auf der Roemerstrasse 1, 76228 Karlsruhe, Germany Page 19 / 40 Phone +49 721 4846-0, Fax +49 721 4846-1019, Email info@pi.de,...
Page 20
5. Loosen the clamping screw of the aperture using the hex key supplied as shown in the figure below. Physik Instrumente (PI ) GmbH & Co. KG, Auf der Roemerstrasse 1, 76228 Karlsruhe, Germany Page 20 / 40 Phone +49 721 4846-0, Fax +49 721 4846-1019, Email info@pi.de,...
Page 21
8. Screw the objective from below into the thread adapter on the objective side of the PIFOC positioner. Do not exceed the maximum torque of 0.6 Nm. Physik Instrumente (PI ) GmbH & Co. KG, Auf der Roemerstrasse 1, 76228 Karlsruhe, Germany Page 21 / 40 Phone +49 721 4846-0, Fax +49 721 4846-1019, Email info@pi.de,...
4: Screw for thread adapter in the objective side Arrows: Direction for mounting the respective thread adapter into the P-725's aperture Physik Instrumente (PI ) GmbH & Co. KG, Auf der Roemerstrasse 1, 76228 Karlsruhe, Germany Page 22 / 40 Phone +49 721 4846-0, Fax +49 721 4846-1019, Email info@pi.de,...
Page 23
7. Screw the objective from below into the thread adapter on the objective side of the PIFOC positioner. Do not exceed the maximum torque of 0.6 Nm. Physik Instrumente (PI ) GmbH & Co. KG, Auf der Roemerstrasse 1, 76228 Karlsruhe, Germany Page 23 / 40 Phone +49 721 4846-0, Fax +49 721 4846-1019, Email info@pi.de,...
Connect the plugs of the P-725 to the corresponding sockets on the electronics (refer to the user manual for the electronics). Physik Instrumente (PI ) GmbH & Co. KG, Auf der Roemerstrasse 1, 76228 Karlsruhe, Germany Page 25 / 40 Phone +49 721 4846-0, Fax +49 721 4846-1019, Email info@pi.de,...
Set the piezo voltage for the controller to 0 V if the P-725 is not used, but should remain switched on to ensure the temperature stability. Physik Instrumente (PI ) GmbH & Co. KG, Auf der Roemerstrasse 1, 76228 Karlsruhe, Germany Page 26 / 40 Phone +49 721 4846-0, Fax +49 721 4846-1019, Email info@pi.de,...
Follow the instructions on starting and operating the P-725 in the user manual for the electronics. Physik Instrumente (PI ) GmbH & Co. KG, Auf der Roemerstrasse 1, 76228 Karlsruhe, Germany Page 27 / 40 Phone +49 721 4846-0, Fax +49 721 4846-1019, Email info@pi.de,...
In open-loop operation: Set the piezo voltage to 0 V on the electronics. Discharging a P-725 not Connected to the Electronics Connect the P-725's voltage plug to the switched-off electronics from PI for 10 seconds. Maintenance NOTICE Misalignment from loosening screws! The P-725 is maintenance-free and precision adjusted.
The current versions of the user manuals can be downloaded at www.pi.ws. Physik Instrumente (PI ) GmbH & Co. KG, Auf der Roemerstrasse 1, 76228 Karlsruhe, Germany Page 30 / 40 Phone +49 721 4846-0, Fax +49 721 4846-1019, Email info@pi.de,...
E-754 E-754 E-754 The resolution of the system is limited only by the noise of the amplifier and the measuring technology because PI piezo nanopositioning systems are free of friction. All specifications based on room temperature (22 °C ±3 °C).
Page 32
* The linearity error of directly driven actuators measured with analog controllers is typically up to 0.1 %. The resolution of the system is limited only by the noise of the amplifier and the measuring technology because PI piezo nanopositioning systems are free of friction.
Note that the decimal places are separated by a comma in the drawings. Fig. 10: P-725.1CDE2, dimensions in mm (please order adapters separately) Physik Instrumente (PI ) GmbH & Co. KG, Auf der Roemerstrasse 1, 76228 Karlsruhe, Germany Page 34 / 40 Phone +49 721 4846-0, Fax +49 721 4846-1019, Email info@pi.de,...
Page 35
Fig. 11: P-725.xCD/.xCL, Dimensions in mm (please order the adapter set separately) Physik Instrumente (PI ) GmbH & Co. KG, Auf der Roemerstrasse 1, 76228 Karlsruhe, Germany Page 35 / 40 Phone +49 721 4846-0, Fax +49 721 4846-1019, Email info@pi.de,...
Page 36
Fig. 12: P-725.xCA, dimensions in mm (please order the adapter set separately) Physik Instrumente (PI ) GmbH & Co. KG, Auf der Roemerstrasse 1, 76228 Karlsruhe, Germany Page 36 / 40 Phone +49 721 4846-0, Fax +49 721 4846-1019, Email info@pi.de,...
In order to fulfill the responsibility as the product manufacturer, Physik Instrumente (PI) GmbH & Co. KG undertakes environmentally correct disposal of all old PI equipment made available on the market after 13 August 2005 without charge. Any old PI equipment can be sent free of charge to the following address: Physik Instrumente (PI) GmbH &...