11.5.6. Screen 4: Sensor Data
Register
x-PEAK C: 
x-PEAK V: 
x-AVG
V:
x-ZERO V:
• PEAK C is the current peak to peak A/D counts
from detector.
• PEAK V is the current peak to peak voltage
from detector.
• AVG V is the running average voltage from 
detector.
• ZERO V is the voltage measured at last zeroing
and the IR sensor output that was stored during
the last purge cycle.
Press the down arrow key to display the fifth screen:
11.5.7. Screen 5: Sensor Data
Register
x-NOISE : 
: 
x-AU
x-AVG AU: 
x-BNCH P:
• NOISE is the largest change in running 
average: 16-point running average of the noise
portion of IR bench output. This reading is
valuable mainly when refrigerant is NOT 
present.
• AU is the current absorbance value.
• AVG AU is the running average of absorbance
level (Average Absorbency). In the absence of
refrigerant, the absorbency is 0.000AU. When
sampling refrigerants, its value varies 
proportionally with the refrigerant concentra-
tion.
• BNCH P is PPM in bench, based on zone gas
selection (uncorrected).
Press the down arrow key to display the sixth screen:
STATUS OPTION Screen
11.5.8. Screen 6: Sensor Data
Register
x-STPPPM: 
x-BNCH L:
x-AMB L :
• STPPPM is PPM corrected to STP (1 atm, 25
deg C).
• BNCH L is micromoles/liter in bench 
(uncorrected).
• AMB L is micromoles/liter corrected to 
ambient pressure.
Press the down arrow key to display the seventh
screen:
11.5.9. Screen 7: Communication
Statistics
NO 
SENT
TRY
1: <ct>
<ct>
2: <ct>
<ct>
3: <ct>
<ct>
• SENT is total messages sent.
• TRY is number of retries needed for a message.
• NO is no response after all four tries.
• 1: <ct> <ct> <ct> is RLDS 1 count for each 
statistic.
• 2: <ct> <ct> <ct> is RLDS 2 count for each 
statistic.
• 3: <ct> <ct> <ct> is RLDS 3 count for each 
statistic.
Press the down arrow key to go back to the 
STATUS OPTION screen. Your changes are saved
and the data is written. Statistics are cleared on reset.
<ct> 
<ct> 
<ct>
Main Screens • 59