Setting The Density Multiplier Switching By External Signals - Toshiba LQ300A00 Series Operation Manual

The insertion type density meter
Table of Contents

Advertisement

10.7.2

Setting the density multiplier switching by external signals

(1) Switching to the setting mode (see Subsection 5.2.5)
First of all, press the [ESC] key of the converter several times (normally once although this varies
with the operation status) to return to the initial menu display. Next, use the [→] key to move the
LCD indicator cursor to the menu number "2" of "2 SETTING MENU" and then press the [SET]
key to display the warning message saying "Test output will be valid.". Make sure that there is no
problem and then press [→] to get into the setting mode. To get into [2: SETTING MENU], it is
necessary to further enter the password "8000". Then, the output will be switched to the
simulated output that is set beforehand.
(2) Selecting the "OTHERS" setting menus (see Subsection 5.2.5)
Initially, the menu list of menu numbers 5 to 8 is displayed. However, by pressing the [UP] key,
this display is switched to the menu list of the next menu numbers 9 to 11. Use the [→] key to
move the LCD indicator cursor to the menu number "11" of "11: OTHERS" and then press the
[SET] key to select the menu of "11: OTHERS."
(3) Selecting of density multiplier switching ON/OFF(see Subsection 5.2.16)
Press the [UP] key once to display "DI: C CHANGE ON/OFF". Press the [SET] key here to
switch to the setup display. Each time the [UP] or [DN] key is pressed, the set value alternates
between "ON" and "OFF". Therefore, when using the function of density multiplier switching by
external signals, set the value to "ON" and then press the [SET] key to verify it. The screen will be
returned to the immediately preceding one; make sure that the value is correctly reset.
(4) Setting density multipliers C2, C3 and C4
Continuing on from (3), press the [UP] key once to switch to the display of "C2:
DensityMultiplier". Switch to the setup display by pressing the [SET] key here and then input the
density multiplier C2. Use the [→] key to move to the digit, and use the [UP] or [DN] key to alter
the numeric value of the relevant digit. When entering input is completed, press the [SET] key to
confirm it. The screen will be returned to the immediately preceding one; make sure that the value
is correctly reset. Continue on to switch to the next setup menu display with the [UP] key and set
C3 and C4 in accordance with the same procedure. When setting the density multiplier C1, refer
to "6.5 Span Calibration".
(5) Returning to the measuring mode
Press the [ESC] key twice to return to the initial menu screen and return to the measuring mode
from the setting mode to resume the normal measurement.