Toshiba LQ300A00 Series Operation Manual page 50

The insertion type density meter
Table of Contents

Advertisement

zG:ZERO RF DATA
DATA: 50.00
[ESC]
[SET] CHANGE
(Previous
[ESC] RETURN
Menu)
[UP]
ma:AVERAGING TIMES
DATA:1
[ESC]
[SET] CHANGE
(Previous
[ESC] RETURN
Menu)
[UP]
dx:
CHANGE RATE LIMIT
DATA:0.00
[ESC]
[SET] CHANGE
(Previous
[ESC] RETURN
Menu)
[UP]
HL:LIMIT TIMES
DATA: 0
[ESC]
[SET] CHANGE
(Previous
[ESC] RETURN
Menu)
[UP]
UR:UPPER RANGE
DATA:3.0
[SET] CHANGE
[ESC]
[ESC] RETURN
(Previous
Menu)
[NOTE]:
For all menus, when data outside the allowed range is
input, the error message is displayed, and the wrong
data is refused. Press any key to return, and input
adequate data.
[DN]
[SET](Set)
[SET](Fix)
[ESC](Cancel)
[DN]
[SET](Set)
[SET](Fix)
[ESC](Cancel)
[DN]
[SET](Set)
[SET](Fix)
[ESC](Cancel)
[DN]
[SET](Set)
[SET](Fix)
[ESC](Cancel)
[DN]
 48 
zG:ZERO RF DATA
RANGE :0.00-100.00 ゚C
DATA:
[SET] SET,[ESC] CANCEL
The zero-point RF data can be set in this menu through manual
input. Normally, the zero-point RF constant is automatically set when
implementing the zero calibration.
ma:AVERAGING TIMES
RANGE : 1-99
DATA :
[SET] SET,[ESC] CANCEL
The moving average times are set. If the output averaged for density
control, etc. is needed, the value from 5 to about 10 is set.
dx:
CHANGE RATE LIMIT
RANGE :0.00-9.99
DATA:
[SET] SET,[ESC] CANCEL
If there is a signal of sharp variation, the change-rate limit function
excludes this signal to restrict a sudden output variation. In this
menu, the permissible variation width is set in terms of "%TS".
HL:LIMIT TIMES
RANGE :0-99
DATA:
[SET] SET,[ESC] CANCEL
The limit times in the change-rate limit function are set. If the variation
width exceeds dx(%TS) that was set in the previous menu, the signal is
excluded for up to the HL times that are set in this menu. At HL=0, the
change-rate limit function does not work.
VALUE OUT OF RANGE
UR:UPPER RANGE
PRESS ANY KEY TO
RETURN