(3-5) Setting the sensitivity and the compound ratio
Continuing on from (3-4), press the [UP] key to switch to the menu of "s0: MAIN OBJ
SENS." and then, after switching to the setup display by pressing the [SET] key, input the
sensitivity of the main object (component 0). Use the [→] key to move to the digit, and use
the [UP] or [DN] key to alter the numeric value of the relevant digit. When entering input is
completed, press the [SET] key to confirm it. The screen will be returned to the
immediately preceding one; make sure that the value is correctly reset. Continue on to
switch to the next setup menu display with the [UP] key. And, in accordance with the same
procedure, set the sensitivities s0 to s5 of components (additives) 1 to 5 as well as
compound ratios R1 to R5 of components (additives) 1 to 5.
(4) Returning to the measuring mode
Press the [ESC] key twice to return to the initial menu screen and return to the measuring mode
from the setting mode to resume the normal measurement.
10.5.5

Simplified Correction on Additives

Previous section describes complete correction on additives in which sensitivities and ratios of mixture
per additive were taken into account in calculation. By using only a sensitivity for the whole mixture (a
mixture of pulp and additives], the additives correction can be simplified as described below. In this
simplified correction, a density multiplier (sensitivity) is obtained separately for each formula and
entered into the list of formula. When a formula is changed, you need to select a new formula number to
change the density mulyiplier to continue measuring density. Up to ten formulas can be entered into the
list.
Procedures for a simplified correction is as follows. See the section 10.5.1 to 10.5.3 for basic
information, and refer to section 10.5.4 Operations for correction.
● Procedures for simplified correction of additives.
Set the density multiplier C to 1.000
Select Calculation Mode for additives correction...No (correction)
Perform zero calibration
Carry out span calibration for each brand, and obtain sensitivity as below.
Sensitivity = (Measured density reading)/(Density obtained by manual analysis)
Enter the sensitivity of a formula into the sensitivity s0 of the component 0 in the formula list.
Select (TOTAL)(total density of a mixture, with additives correction] for the calculation mode
Select a brand list number that corresponds with the mixture to be measured.
Run the fluid to measure and measure density
(Note1) Set all sensitivities and ratios to 0.00 and 0.000 respectively for all of the components 1 to 5.
(Or to the default values set at shipping.)
See "6.4 Zero calibration"