Toshiba LQ300A00 Series Operation Manual page 49

The insertion type density meter
Table of Contents

Advertisement

dt:DELAYED SYNC.TIME
DATA:0.5
[ESC]
[SET] CHANGE
(Previous
[ESC] RETURN
Menu)
[UP]
zp:ZERO POINT PHASE
DATA:80.00
[SET] CHANGE
[ESC]
[ESC] RETURN
(Previous
Menu)
[UP]
zT:ZERO LIQUID TEMP.
DATA:25.00
[ESC]
[SET] CHANGE
(Previous
[ESC] RETURN
Menu)
[UP]
cG:RF COEF.
DATA: 0.00
[SET] CHANGE
[ESC]
(Previous
[ESC] RETURN
Menu)
[UP]
[DN]
[SET](Set)
[SET](Fix)
[ESC](Cancel)
In external synchronized operations, the delayed time from when the
external contact input is turned ON until the measurement starts is
[DN]
set. For details, refer to Section 6.7.
[SET](Set)
[SET](Fix)
[ESC](Cancel)
[DN]
[SET](Set)
[SET](Fix)
[ESC](Cancel)
[DN]
[SET](Set)
[SET](Fix)
[ESC](Cancel)
 47 
dt:DELAYED SYNC.TIME
RANGE :0.1-99.9 min
DATA :
[SET] SET,[ESC] CANCEL
zp:ZERO POINT PHASE
RANGE :0.00-359.99
DATA :
[SET] SET,[ESC] CANCEL
The zero-point phase can be set in this menu through manual input.
Normally, the zero-point phase is automatically set when
implementing the zero calibration.
zT:ZERO LIQUID TEMP.
RANGE :0.00-100.00 ゚C
DATA :
[SET] SET,[ESC] CANCEL
The zero-point liquid temperature can be set in this menu through
manual input. Normally, the zero-point liquid temperature is
automatically set when implementing the zero calibration.
cG:RF COEF.
RANGE : -9.99-9.99
DATA:
[SET] SET,[ESC] CANCEL
The RF correction factor used for density calculation is set. Before
shipment, this value is already set in accordance with the product's
characteristics in the shipping test. If you change this value, it will affect
the performance; therefore, normally, do not change it.