5.2.11

Zero calibration display and operating procedures

5:SET PARAMETERS
6:ZERO CALIBRATION
7:SPAN CALIBRATION
8:ANGLE ROTATION
Move the cursor to "6" with [→] key, and press [SET] key to select [6:ZERO CALIBRATION]
p PHASE :81.23 ゜ N=0
z
zT TEMP. :26.55 °C
[ESC]
zA TEMP. :26.55 °C
(Previous
*X DENSITY : 0.05 %TS
Menu)
5.2.12

Span calibration display and operating procedures

5:
SET PARAMETERS
6:ZERO CALIBRATION
7:SPAN CALIBRATION
8:ANGLE ROTATION
[ESC]
(Previous
Menu)
[SET](Set)
New data displayed
Move the cursor to "7" with [→] key, and press [SET] key
to select [7:SPAN CALIBRATION]
C1:DensityMultiplier
DATA:1.000
[SET] CHANGE
[ESC] RETURN
 49 
ZERO POINT
CALIBRATION
HAS BEEN COMPLETED
Press the [SET] key for zero calibration. In zero calibration, the
phase, the liquid temperature, the atmospheric temperature, and
RF data of zero point will be replaced with the present measured
values. And also, angle rotation "N" will be set to zero.
Note:The display will return to the left display automatically after
approximately one second.
[SET](Set)
C1:
DensityMultiplier
RANGE :0.000-9.999
[SET](Fix)
DATA :
[ESC](Cancel)
[SET] SET,[ESC]CANCEL
The density multiplier C1 should be set to the
suitable value so that the measured density value is
corresponding to the manual analysis value. For
details, refer to Section 6.5.