Replacing The Backup Battery; Replacing/Relocating Detectors; Annual System Check - Visonic PowerMaster 360 Installer's Manual

Table of Contents

Advertisement

Fault
What it means
commander is near the end of its useful life.
MISSING
A device or detector has not reported for
some time to the control panel.
NOT NETWORKED
A device was not installed or not installed
correctly, or, cannot establish communication
with the control panel after installation.
RSSI LOW
The GSM communicator has detected that
GSM network signal is weak
SIREN AC
There is no power to the siren
FAILURE
TAMPER OPEN
The sensor has an open tamper
TROUBLE
The sensor reports trouble
SOAK TEST FAIL
Detector alarms when in Soak Test mode

6.2 Replacing the Backup Battery

Replacement and first-time insertion of battery pack is similar, see Figure 3.2.
With a fresh battery pack, correct insertion and tightened battery compartment lid, the TROUBLE indicator should
extinguish. However, the "MEMORY" message will now blink in the Virtual Keypad display (caused by the "tamper"
alarm you triggered when opening the battery compartment lid). Clear it by arming the system and immediately
disarming.

6.3 Replacing/Relocating Detectors

Whenever maintenance work involves replacement or re-location of detectors, always perform a full diagnostic test
according to section 4.8.
Remember! A "poor" signal is not acceptable.

6.4 Annual System Check

Note: The PowerMaster 360 system must be checked by a qualified technician at least once every three (3) years
(preferably every year).
The annual system check is designed to ensure proper operation of the alarm system by performing the following checks:
Periodic test
Arm/disarm function
No trouble messages are displayed on the Virtual Keypad
The clock displays the correct time
Reporting: generating an event to be transmitted to the Monitoring Station and to the user.
D-305735 PowerMaster 360 Installer's Guide
6. MAINTENANCE
Possible Solution
is available and connected to the device.
Replace the device battery.
• Make sure the device is physically present.
• Check the display for device faults, for
example, low battery.
• Use RF diagnostics to check the current
signal strength and during the last 24
hours.
• Replace the battery.
Replace the device.
• Make sure the device is physically present.
• Use RF diagnostics to check the current
signal strength and during the last 24
hours.
• Open the device cover and replace the
battery or press the tamper switch.
Enroll the device again.
Move the Panel and GSM unit to another
location.
Make sure that the AC power supply is
connected properly
Close sensor tamper
Replace the sensor
If you wish to continue the Soak Test, no
further action should be taken.
If you wish to abort the Soak Test, disable the
Soak Test (see section 4.4.6).
55

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents