General requirements

WARNING!
Do not open the transmitter, sensor, or junction box enclosure when in a classified area or when an
explosive atmosphere may be present unless the power to the transmitter has been removed.
When connecting cable wires, use a small screwdriver to gently press down and hold the
spring connector open. Insert the appropriate wire into the open connector hole, releasing
the screwdriver to secure the wire as shown in
The connector will accommodate wire sizes between 14 and 20 AWG.
Rotate the PCB assembly to the desired position and line up the standoffs with the
mounting holes.
Insert standoffs in the appropriate horizontal or vertical mounting holes.
Tighten standoffs to secure the PCB assembly.
Replace faceplate and tighten Pull Here knobs.
Return power to the detector and replace the housing cover.
Figure 2-9:
PCB Assembly Rotated
Figure
2-10.
Installation
13