11.1.1.3

Description of display settings of EasyOne Air

Setting
Purpose
Language
To set the language
Brightness
To set the brightness of the touchscreen
Screensaver after
To set the time, after which the
screensaver is enabled
Power-off after
To set the time, after which EasyOne Air
is powered off to save battery
Volume/Time curve
To enable the display of the volume-
time curve in addition to the flow-
volume curve in the reports
Show predicted range
To enable visualization of the maximum
curve and minimum curve in the
volume-time curve and flow-volume
curve in the reports
 Description of display settings
Operator's Manual V1.1 • EasyOne Air • EasyOne Connect
© ndd Medizintechnik AG • Technoparkstrasse 1, 8005 Zurich, Switzerland • www.ndd.ch
Ethnic correction
Some studies for predicted values take into account the differences between
certain ethnic groups. However, most studies that are used for spirometry
were conducted on Caucasian subjects and are therefore most appropriate for
use with Caucasian patients.
When entering patient information, you are presented with a list of options for
four ethnic groups. In this setting, you can define the corrections for these
ethnic groups. The adjustment is made to the values for Caucasian patients.
There is an exception to this function. If specific values are available for the
chosen group of predicted values and the chosen ethnic group, they will be
used instead of the correction entered here.
The American Thoracic Society's publication, Lung Function Testing: Selection
of Reference Values and Interpretative Strategies [8], provides guidance on
the subject of ethnic correction. This paper recommends using 88% of the
Caucasian values when testing African patients and provides general
guidance in selecting adjustments for other ethnic groups.
List of bibliographic references, 132
>
>
On this screen, you can choose settings that are related to the displayed
software and to the touchscreen.
Options
English, French, German,
Spanish
1% – 100%
5 s – 600 s
20 s – 1200 s
Selected, deselected
Selected, deselected
Default setting
(Determined by
regional setting at
first-time setup)
50%
120 s
600 s
Selected
Selected
99/144