Performing spirometry
7.6.2
Performing a complete test using EasyOne Connect together with EasyOne Air
Operator's Manual V1.1 • EasyOne Air • EasyOne Connect
© ndd Medizintechnik AG • Technoparkstrasse 1, 8005 Zurich, Switzerland • www.ndd.ch
Required materials
A new EasyOne Flow Tube for each patient
For an FVL test, an MVV test, or an SVC test, a nose clip is required.
Prerequisites
You have connected EasyOne Air to EasyOne Connect.
You have added the patient data to the database of EasyOne Connect.
Or the patient data is already in the database of EasyOne Connect.
You are wearing disposable gloves.
You have instructed the patient on how to perform the test before you
start with the test.
Connecting EasyOne Air to EasyOne Connect, 63
Adding patient data to the database of EasyOne Connect, 86
About preparation of spirometry and instructions for the patient, 52
WARNING
PATIENT CROSS-CONTAMINATION
If you reuse the EasyOne Flow Tube, contamination from an infected patient
can be deposited on the EasyOne Flow Tube. The contamination from an in-
fected patient can later be passed on to the next patient who is tested.
You cannot clean or disinfect the EasyOne Flow Tube in any way. The
EasyOne Flow Tube is for single-patient use only.
Always replace the EasyOne Flow Tube with a new one between patients or
when performing spirometry on yourself.
Only use an original EasyOne Flow Tube by the manufacturer.
PATIENT HEALTH HAZARD
Performing spirometry can cause an asthma attack or bronchospasm.
If the patient is on medication, check the contraindications for the medica-
tion.
CAUTION
PATIENT HEALTH HAZARD
Pulmonary function tests require maximum effort on the part of the patient and
may lead to dizziness.
Make sure the patient cannot be injured by objects in the vicinity if the patient
falls.
Watch the patient for signs of dizziness and support the patient if necessary.
Do not leave the patient unattended during a test.
The EasyOne Flow Tube is a single-patient-use disposable and hygienically
packaged. Visually check the wrapper of a new EasyOne Flow Tube for defects.