Utilities > Configuration > Test > Svc; Utilities > Configuration > Test > Mvv; Utilities > Configuration > Test > Calcheck - NDD EasyOne Air Operator's Manual

Hide thumbs Also See for EasyOne Air:
Table of Contents

Advertisement

Configuration
11.2.4.5
Utilities > Configuration > Test > SVC
11.2.4.6
Utilities > Configuration > Test > MVV
11.2.4.7
Utilities > Configuration > Test > CalCheck
Operator's Manual V1.1 • EasyOne Air • EasyOne Connect
© ndd Medizintechnik AG • Technoparkstrasse 1, 8005 Zurich, Switzerland • www.ndd.ch
Print Lung Age
You can choose to print the lung age on reports. When the calculated lung age
is lower than the patient's actual age, the patient's actual age is displayed.
To display the lung age only for smokers, choose
down list. You can set a patient's smoker status when you add patient data to
the database or edit patient data.
To always display the lung age, choose On.
To never display the lung age, choose Off.
The default configuration is
Parameter
To select which diagnostic parameters to include on the test result screens
and on the printed reports, choose
Type
To make regular breathing mandatory before the breathing maneuver, choose
Tidal Breathing
Required.
Parameter
To select which diagnostic parameters to include on the test result screens
and on the printed reports, choose
Syringe Volume
Calibration syringes are available in various sizes. You can choose the volume
of the calibration syringe from the drop-down list.
Type Selection
One test with three trials is required for the Single Flow calibration check. For
the Multi Flow calibration check, three tests comprising three trials, each with
different flow rates, are required.
Reported Graph
You can select which type of curve you prefer for the calibration check report.
Smoker only
Smoker
only.
Select FVC
and
Select
Select FVC
and
Select
from the drop-
FVL.
FVL.
117
114/144