To Set The Memory Depth - Rigol DS4024E User Manual

Ds4000e series; digital oscilloscope
Hide thumbs Also See for DS4024E:
Table of Contents

Advertisement

RIGOL

To Set the Memory Depth

The memory depth refers to the number of waveform points that the oscilloscope
can store in a single trigger sample and it reflects the storage ability of the
acquisition memory. DS4000E provides up to 14 Mpts memory depth for each
channel.
The relation of the memory depth, sample rate and waveform length fulfills the
equation below:
Memory Depth = Sample Rate × Waveform Length
Wherein, Waveform Length = Horizontal Time Base × number of grids horizontally
on the screen. For DS4000E, the number of grids horizontally on the screen is always
14. Therefore, with the same horizontal time base, higher memory depth can ensure
higher sample rate.
Press Acquire  Mem Depth, turn the multi-function knob
desired memory depth (the default is auto); then, press the knob. You can also press
Mem Depth continuously to switch the current memory depth.
The memory depth can be set to Auto, 7 kPoints, 70 kPoints, 700 kPoints and 7
MPoints.
In "Auto" mode, the oscilloscope selects the memory depth automatically
according to the current sample rate. As the sample rate might change when
adjusting the horizontal time base, the memory depth might also change
accordingly.
4-6
Trigger Point
T
Pre-sample
Memory Depth
Chapter 4 Sample System Setting
Delayed Sample
to select the
DS4000E User's Guide