The trigger coupling setting determines the type of signal components that will be
transmitted to the trigger circuit.
Press MENU (in the trigger control area (TRIGGER) at the front panel)  Setting
 Coupling and turn the multi-function knob
type (the default is DC). You can also press Coupling continuously to switch the
current coupling mode.
DC: allow DC and AC components into the trigger path.
AC: block the DC components and attenuate signals lower than 8 Hz.
LF Reject: block the DC components and reject the low-frequency components
lower than 5 kHz.
HF Reject: reject the high-frequency components higher than 50 kHz.
Note: Trigger coupling has no relation with channel coupling (refer to "To Specify
the Channel Coupling").