Thermo Scientific Orbitrap Fusion Lumos Getting Started Manual page 37

Orbitrap tribrid series
Hide thumbs Also See for Orbitrap Fusion Lumos:
Table of Contents

Advertisement

To set up the syringe and syringe pump
(Not shown) Load a clean, 500 μL
1
syringe with the sample solution.
(Syringe pump, top view)
4
Squeeze the syringe pump's
release button and slowly move
the pusher block until it contacts
the syringe plunger.
Thermo Scientific
Using a two-piece, fingertight fitting, connect 4 cm (1.5 in.)
2
of Teflon™ tubing to the (black) LC union.
(LC union connection,
cross-sectional view)
3
Hold the plunger of the syringe in place and carefully insert the tip of the
syringe needle into the free end of the Teflon tubing. Then, place the syringe
into the syringe pump with the end flange against the pump.
Note If necessary, use the syringe needle tip to enlarge the opening slightly
in the end of the Teflon tubing.
3
Connecting the Inlet Plumbing
Setting Up the Syringe and Syringe Pump
Orbitrap Tribrid Series Getting Started Guide
19