Table 2. Guidelines for setting operating parameters for LC/H-ESI/MS
LC flow
Spray
rate (μL/min)
voltage (V)
Up to 15
Pos: 3500
Neg: –2500
16–99
Pos: 3500
Neg: –2500
100–199
Pos: 3500
Neg: –2500
200–500
Pos: 3500
Neg: –2500
a
Positive and negative polarity modes
Table 3. Guidelines for setting operating parameters for LC/APCI/MS
LC flow rate
(μL/min)
Sheath
Up to 5
15
5–199
25
200–500
45
a
Positive and negative polarity modes
Table 4. Guidelines for setting operating parameters for LC/NSI/MS
Spray voltage (V)
Positive mode: 1200
Negative mode: –600
Thermo Scientific
Gas (arbitrary units)
a
Sheath
Auxiliary
5
2
25
5
35
7
50
10
Gas (arbitrary units)
Auxiliary
Sweep
5
5
5
Sweep gas (arbitrary units)
2
Ion transfer tube
temp (°C)
Sweep
Fusion MS: 275
0
Lumos MS: 325
Fusion MS: 275
0
Lumos MS: 325
Fusion MS: 300
0
Lumos MS: 325
1
325
Ion transfer tube
temp (°C)
Fusion MS: 250
0
Lumos MS: 300
Fusion MS: 250
0
Lumos MS: 300
Fusion MS: 275
0
Lumos MS: 325
LC Flow Rate Ranges
Vaporizer
Typical nitrogen
temp (°C)
gas consumption
20
75
200–275
275–350
Vaporizer
Corona
temp (°C)
discharge
current (μA)
275
Pos: 4
Neg: –10
325
Pos: 4
Neg: –10
350
Pos: 4
Neg: –10
Ion transfer tube temperature (°C)
275
Orbitrap Tribrid Series Getting Started Guide
1
Introduction
(L/min)
3.5
8.0
10.5
15.0
a
5